JSR Ultrasonics PureView Series Скачать руководство пользователя страница 17

PureView™ Pulser-Receiver Operator Manual 

 

 

  Page 17   

Imaginant Inc. 

The PureView red status indicator ‘Pulse’ LED will illuminate to indicate that the pulser is being 
triggered.    Enable the pulser power supply to produce transducer excitation pulses on the T/R 
connector.    Adjust the frequency of pulser firing such that all echoes from one excitation pulse 
have subsided before a new excitation pulse is generated. 

Adjust the Damping and Energy controls to obtain the desired excitation pulse characteristics. 

Adjust the receiver gain control so that signals of interest (signals from the PureView Output 
connector) do not exceed the range of +500mV to - 500mV into a 50

 termination. 

Adjust the high and low pass filter cutoff frequencies (as desired). The High pass filters can be used 
to eliminate low-frequency energy from signals of interest. The low pass filter can be used to 
reduce the receiver bandwidth and improve the signal to noise ratio. 

 

IMORTANT NOTE - The PureView receiver contains circuitry to protect its receiver from the large transducer 
excitation pulse.    The operation of this protection circuit causes two short pulses to appear in the receiver 
Output signal, and these pulses bracket the time when the pulser ‘fires’.    The first short pulse occurs 
approximately 70ns after the rising edge of the Trig / Sync signal.    The second short pulse follows the first by 
300ns or more (depending on the PureView model).     

The Receiver Output signal will not provide any ultrasound signals from its T/R or Through inputs during the 
interval of time between these two short pulses.    During this interval the Receiver Output waveform contains 
no meaningful signal content.    Although the two short pulses may look similar to ultrasound signals, neither 
of these short pulses is representative of any signals produced by the transducer(s) during that interval.     

An example of the receiver-protection pulses is shown below.    This is the output of a PureView unit where 
the horizontal time scale is 100ns / major division.    The purple trace is the PureView Trig / Sync pulse.    The   
blue trace is the Receiver Output signal with the Receiver gain set to 20dB.    The two short pulses in the (blue) 
Receiver Output waveform follow the rising edge of the (purple) Trig / Sync pulse by 70ns and 640ns 
respectively.    The exact timing of these pulses is PureView model dependent and the second pulse may 
appear earlier or later in time than what is shown, depending on the model.    In some low frequency 
PureView models, the second pulse may occur up to 1.2us after the rising edge of the Trig / Sync signal. 

 

Содержание PureView Series

Страница 1: ...PureView Series USB Pulser Receiver Operator Manual PureView User Manual Version 1 3 January 2021 Copyright 2021 All Rights Reserved...

Страница 2: ...poration in the United States and or other countries Safety Advisories English The PureView system is designed for manufacturing laboratory and office use Fran ais French Le syst me PureView est con u...

Страница 3: ...dirigido la seguridad total podr a potencialmente ser comprometida Deutsch German Verwenden Sie die PureView Instrumentenausr stung wie durch den Hersteller verwiesen Wenn Sie verwenden nicht wie ver...

Страница 4: ...OPERATION 9 PUREVIEW INSTRUMENT FUNCTIONAL BLOCKS 9 Control Logic and USB Interface 10 Pulser Energy and Damping 10 Pulser Voltage 10 T R and Through Select 11 Amplifier Attenuator 11 Low Pass Filter...

Страница 5: ...PULSER CONTROLS IN JSR NET CONTROL PANEL 19 Trigger Enable 19 Trigger Source 19 PRF 19 Pulser Voltage Supply Enable 20 Pulser Voltage 20 Pulser Energy 20 Damping 20 Trigger Zin 20 INSTRUMENT STATUS I...

Страница 6: ...PureView Pulser Receiver Operator Manual Page 6 Imaginant Inc Miscellaneous 23 Environmental 23...

Страница 7: ...ber they must ship the defective product to Imaginant The RMA number must be clearly marked on the outside of the shipping package The customer is responsible for the costs associated with shipping pa...

Страница 8: ...to Mini B Additional Equipment Required Customer Supplied 1 Computer running Windows O S with an available USB port preferably an enhanced power USB port 2 Ultrasonic Transducer s 3 Trigger signal sou...

Страница 9: ...s water resistant although care should be taken to keep the instrument dry Connections are easily made to PureView hardware by attaching coaxial cables to PureView s four female SMA interface connecto...

Страница 10: ...parameters such as pulse duration pulse energy and pulse amplitude Note that a change to either the Energy or the Damping value will affect all three pulse parameters The pulser produces constant amp...

Страница 11: ...ides control over the PureView receiver s gain The receiver gain can be set in 0 5dB increments across the gain range of 11 5dB to 70db Low Pass Filters The low pass filter enables a user to reduce th...

Страница 12: ...the T R connector to excite the transducer The transducer converts energy from the electrical excitation pulse into an ultrasonic pulse that is propagated into a test material Ultrasonic return signal...

Страница 13: ...oaxial cable The PureView Receiver can provide a linear output signal of 500mV maximum into a 50 load and signals of interest should not be amplified beyond these limits Over amplification of signals...

Страница 14: ...ED indicators that illuminate to indicate the instrument status as shown in the image below The Power and Pulse LED indicators indicate the following Indicator Function Power The Blue LED indicates th...

Страница 15: ...e blue power LED indicator on PureView should illuminate The first time that a PureView unit is connected to the PC there may be a slight delay as the operating system finds and loads the hardware dri...

Страница 16: ...ing figure Outpu t Trig Sync Through T R Sig Trig Oscilloscope Transmitting Transducer Receiving Transducer Figure 3 Through Mode Operation PureView Operation The following sequence of actions describ...

Страница 17: ...pulses bracket the time when the pulser fires The first short pulse occurs approximately 70ns after the rising edge of the Trig Sync signal The second short pulse follows the first by 300ns or more de...

Страница 18: ...an example the image below shows the output signal of a PureView unit with the receiver gain set to 70dB The horizontal time scale is 500ns major division The action of the receiver protection circui...

Страница 19: ...odel the high pass filters may have values of 1MHz 5MHz 30MHz 65MHz or other values Pulser Controls in JSR Net Control Panel Trigger Enable This control allows a user to enable or disable the Pulser f...

Страница 20: ...connector to either 50 or High Z 18k Selecting a value of 50 connects an internal 50 terminating impedance to the Trig Sync connector which is appropriate for terminating a 50 coaxial cable You may se...

Страница 21: ...a return is recommended the Representative will ask for details regarding the repair issue and will issue an RMA number If possible when returning equipment to Imaginant please package the equipment...

Страница 22: ...Trigger Source Internal or External user selectable rising edge triggers Pulser External Trigger Input Signal 1V to 5V positive going pulse Sync Signal Output Pulse Rising edge transition 3ns Width 60...

Страница 23: ...ity low loss USB cable should be employed to avoid any voltage drop through the cable Physical Weight 0 152kg 0 33 lbs Enclosure Dimensions 124 5mm x 50 3mm x 25 4mm 4 9 L x 1 98 W x 1 H Length includ...

Отзывы: