background image

©JRI 

 

- The recorder’s power supply is self-protected, but is not equipped with a power cutting system. A 
power-breaking device should be included on the power supply line (switch or circuit breaker). 
The device should be placed near the recorder, easily accessible and marked as being the recorder’s 
power breaker. 

T

HE

 

MECHANISMS POWERED BY THE MAINS ARE NOT EXTRACTABLE

 

IV.

 

TECHNICAL PROPERTIES 

a)

 

Thermograph 

Description 

Self-contained temperature probe – recorder 

Number of measurement 
channels 

1 or 2 

Measuring range 

(From -100 to +250°C) 

0-100°C, -100+50°C, -90+10°C, -35+15°C, -5+25°C, -
25+25°C, 
-10+40°C, 0+120°C, +30+130°C 

Connection  Capillary tube length: standard 3 m (from 1 to 

20m) 

Weight  Approximately  1.8kg 

 

b)

 

Thermograph features 

 

Accuracy  ± 2 % of measuring range at 23° C 

Number of measuring channels  1 or 2 

Full stroke  45 mm 

Viewing  Diagram disk Ø 125 mm 

Length of recording (period)  1R / 7 days, 1 R / 24 hours 

Power supply, self-contained  Battery - LR 6 - 1.5 V, 1 year (AA) 

Appointed operating conditions  -15° C + 60° C 

Storage conditions  -35° C + 65° C 

Size  144 x 144 x 119 mm 

Degree of protection  IP 20 

Accessories :

       Diagrams (per 100)  Dxxxxx (see appliance’s label) 

Fibre point pens (per 5)  S97601 (01 violet, 02 black, 03 red) 

Fibre point pen’s capacity  About 90 m 

Optional :

       

Clockwork mechanism powered by an 

external power source 

230 V AC 50 / 60 Hz 0.015 A 
24 V AC 

 

 