Troubleshooting

Problem

Solution

The ambient temperature
is not within the range 
between 5

 and 40

(41

~

104

).

Place the 
in a room for at least 30
minutes at room 
temperature between 
5

 and 40

(41

~

104

)

thermometer

The thermometer is not 
functioning properly.

Unload the battery, 
wait for 1 minute and 
repower it. If the 
message reappears, 
contact the retailer 
for service.
 

Error message              

20

19

21

22

Specifications

Forehead mode: 34.0

~43.0

(93.2

~109.4

)

Object mode: 0

~100

(32

~212

)

Forehead mode: 

±

0.2

 (0.4

) during 35.5

~42.0

 (95.9

~107.6

at 15

~35

 (59.0

~95.0

) operating temperature range

±

0.3

 (0.5

) for other measuring and operating 

temperature range
Object mode: 

±

4% or 

±

2

(4

) whichever is greater

Approx.

grams(with batteries)

118

DC3V(2

×

AA

 

battery)

Measuring range

Laboratory accuracy

Display resolution

0.1

 or 0.1

Measure time

Operating temperature 
range:

Storage and transport 
temperature range

-25

~ 55

 (-13

~131

), 15%~95%RH, non-condensing

70kPa ~ 106kPa

Atmospheric Pressure : 

Clinical accuracy

Dimension

Weight

Battery

Battery life

Approx. 3000 readings 

Measuring site    

        

Forehead(Forehead Mode)

 Oral (This thermometer converts the forehead temperature
 to display its “oral equivalent.”)

Reference body site

Operation mode                  Forehead mode(Adjust mode)

Ingress protecting rating

         
    

 IP22

Shock

withstands drop of 3 feet

Expected service life

Three years 

5

~40

(41

~104

), 15%~85%RH, non-condensing

70kPa ~ 106kPa

Atmospheric Pressure : 

Battery Replacement

Contraindication

         
    

 No contraindication

Troubleshooting

Care And Cleaning

1.  The probe window must be kept clean, dry, and undamaged at all
     times to ensure accurate readings. The accuracy of temperature 
     readings can be affected by damage to the probe window, or the 
     presence of dirt, fingerprints, dust and other soiling 
     compounds on the probe window. Degraded sensors can degrade
     performance or cause other problems.

 

2

For cleaning

       

1) Soak a clean soft cloth in drinking water,wring it out,and then 

          wipe the thermometer(including probe) no less than 3 times

       

2) Visual or use magnifying glass to observe the thermometer have

          no visible dirt and stains ,then use another clean soft cloth to 
          wipe the thermometer residue water

       

3) Put the thermometer in the original packaging.

  3.For disinfection

       

1) Soak a clean soft cloth in drinking water,wring it out,and then 

          wipe the thermometer(including probe) no less than 3 times

       

2) Visual or use magnifying glass to observe the thermometer have 

          no visible dirt and stains ,then use another clean soft cloth to 
          wipe the thermometer residue water

       

3) Using a clean soft cloth dipped in 70% medical alcohol,wipe the 

           probe for 3 Times,each time 1 minute.
      4) Using a clean cotton swab dipped in 70% medical alcohol,wipe 
           the sensor window 3 times

       

5) Wait at least 10 minutes to let the alcohol volatilization and put it 

           in the original packaging.  
4.  Do not put the thermometer into water directly.
5.  Store the thermometer in a dry location, free from dust and 
     contamination and away from direct sunlight.
6.  

Put the thermometer back to the original packaging after using. 

1.  

 right corner  of  LCD display as

 

shown

 

   

in

 

Figure 8.

When         flashes, it indicates that the power is low, but you can 

     continue to measure; replace battery when "         " 

 in the 

 

remains 

     displayed

lower

2.  Slide battery cover down as shown in Figure 9.

3.  Remove battery and install 2 new 

 alkaline batteries 

AA

     as shown 

 . 

in Figure 10

4.  Slide battery cover back on.

Figure 8

Figure 9

Figure 10

Approximately 1 second

140.0*37.8*34.8mm

0-1 year:
Clinical bias: -0.17

℃( -

0.31

℉) ;

Clinical repeatability: 

0.06

℃(

0.11

℉) ;

Limits of agreement: 0.24

℃(

0.43

℉)

1

-

5

 

years:

Clinical bias: -0.17

℃( -

0.31

℉) ;

Clinical repeatability: 

0.07

℃(

0.13

℉) ;

Limits of agreement: 0.22

℃(

0.40

℉)

over

 

5

 

years

:

Clinical bias: -0.17

℃( -

0.31

℉) ;

Clinical repeatability: 

0.20

℃(

0.36

℉) ;

Limits of agreement: 0.26

℃(

0.47

℉)

6

Beep Alarm

If

 

thermometer

 

detects

 

a

 

body

 

temperature

 ≥ 

37

.

8

(

100

.

0

under

 

forehead

 

mode

there

 

will

 

be

 

a

 

long

 

beep

 

sound

 

followed

 

by

three

 

short

 

beep

 

sound

 

to

 

warn

 

the

 

user

.

语音款没有

beep,

台湾定制蜂鸣和振动共存,在蜂鸣关了之后振动打开,常规款有语音的不能有这个发烧报警

Содержание DET-3015b

Страница 1: ...light emitted from the forehead It converts the measured heat into a temperature reading displayed on the LCD When properly used it will quickly assess your temperature in an accurate manner Indicatio...

Страница 2: ...the Illustration For Use Voice The thermometer will broadcast the result after finishing measurement Language Switch Please see the Real Time Clock Setting to learn how to change between Chinese and E...

Страница 3: ...shing automatically Setting Press and release the to advance one hour until the correct hour appears After the hour is set press Button the Minute figure is flashing automatically Setting Button Set t...

Страница 4: ...ly 30 minutes prior to use 12 It is not intended for use in the oxygen rich environment and presence of flammable anesthetic mixture with air oxygen or nitrous oxide To measure object temperature 1 yo...

Страница 5: ...dest reading stored in memory See Figure 7 Each memory also records the measurement date time mode icon Bluetooth requirements The thermometer requires a bluetooth device with Bluetooth 5 0 or later A...

Страница 6: ...d sensors can degrade performance or cause other problems 2 For cleaning 1 Soak a clean soft cloth in drinking water wring it out and then wipe the thermometer including probe no less than 3 times 2 V...

Страница 7: ...the manufacturer s representative for assistance if needed in setting up using or maintaining the thermometer or to report unexpected operation or events Troubleshooting Symbol Explanation Caution 95...

Страница 8: ...arty responsible for compliance could void the user authority to operate the equipment Note This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to...

Страница 9: ...Hz are intended to decrease the likelihood that mobile portable communications equipment could cause interference if it is inadvertently brought into patient areas For this reason an additional factor...

Страница 10: ...immunity of this equipment and result in improper operation Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm...

Отзывы: