37
36
Electromagnetic Compatibility Information
Electromagnetic Compatibility Information
WARNINGS!
· This device should not be used in the vicinity or on the top of other
electronic equipment such as cell phone, transceiver or radio
control products. If you have to do so, the device should be
observed to verify normal operation.
· The use of accessories and power cord other than those specified,
with the exception of cables sold by the manufacturer of the
equipment or system as replacement parts for internal components,
may result in increased emissions or decreased immunity of the
equipment or system.
· Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation.
· Use of accessories, transducers and cables other than those
specified
or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased
electromagnetic
immunity of this equipment and result in improper operation.
· Portable RF communications equipment (including peripherals
such
as antenna cables and external antennas) should be used
no closer
than 30 cm (12 inches) to any part of the device,including cables
specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
· Portable and mobile RF communications can affect the device.
The device needs special pre-cautions regarding EMC according to
the EMC information provided in the accompany documents.
· Do not use the devices in the MR environment.
· The Operator should not use the system and should inform the
customer service, if the ESSENTIAL PERFORMANCE is lost or
degraded due to EM DISTURBANCES.
· PRECAUTION: The performance of the device may be degraded
should one or more of the following occur:
- Operation outside the manufacturer’s stated temperature and
humidity range.
- Storage outside the manufacturer’s stated temperature and
humidity range.
- Mechanical shock (for example, drop test) or degraded sensor.
- Patient temperature is below ambient temperature.
Electromagnetic Compatibility Information
35
6
Beep Alarm
If
thermometer
detects
a
body
temperature
≥
37
.
8
℃
(
100
.
0
℉
)
under
forehead
mode
,
there
will
be
a
long
beep
sound
followed
by
three
short
beep
sound
to
warn
the
user
.
语音款没有
beep,
台湾定制蜂鸣和振动共存,在蜂鸣关了之后振动打开,常规款有语音的不能有这个发烧报警