Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The JC-1321 [KD-558] is intended for use in the electromagnetic environment specified below.
The customer or the user of the JC-1321 [KD-558] should assure that it is used in such an
environment.
IMMUNITY test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of
the JC-1321 [KD-558], including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
80 MHz to 800 MHz
P
d
2
.
1
=
800 MHz to 2,5 GHz
P
d
3
.
2
=
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment
marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Содержание JC-1321
Страница 50: ...50 03 00 1 1 5 1 MEM 2 START 3 4 LCD 5 6 7 8 9 10 11 12 13 22 30 LCD 2x60 IEC 60601...
Страница 52: ...52 300 mm Hg 15 mm Hg 3 7 1 2 AV 3 4 5 8 9 10 11 1 CV 25 2 0 14s 53 12 13 14 15 B 15 FCC...
Страница 54: ...54 3 4 1 2 1 2 3 5 4 200 5 6 START 6 U1 U2 6 1...
Страница 56: ...56 7 3 7 4 7 5 7 3 7 4 7 5 MEM 7 6 7 7 7 8 MEM 7 9 MEM 7 6 7 7 7 8 7 9 1 START 8 8 MEM 8 START MEM 9...
Страница 58: ...58 Er 3 Er 4 Er 5 300 mm Hg 5 Er 6 3 15 mm Hg Er 7 EEPROM Er 8 Er A 5 1 2 3 4 5 6 2 7 8 9 10...