Jorvet J1550 Скачать руководство пользователя страница 9

11. Alarm Removal Key

Alarm and Processing Control: when the O

2

 concentration is higher than the set point and remains 

constant for some time (to avoid false alarm for an abrupt increase of partial concentration), the O

2

 

Concentration Indicator Lamp is flashing with the alarm. Start both the left fan and the right fan to 
enhance ventilation inside and outside. If the O

concentration remains unchanged, open the ICU door 

to enhance ventilation. When the O

2

 

concentration is equal to or lower than the set point, switch off the 

O

2

 indicator lamp and close the alarm. If fans are not needed when other functions are working the 

fans automatically stop.

1. Disinfection & Sterilization: after incubation or caring of one pet finishes treatment and care, it is
suggested to disinfect and sterilize this machine. Do this once every week when not in use.
2. Separate cleaning of the Humidifier
If the Humidifier fails to produce moisture, or the water spills out, take the following measures:
a) As shown in Figure 8.2.1-a,open the cover plate of the right Control Rack.
b) As shown in Figure 8.2.1-b and Figure 8.2.1-c,remove Water Level Sensors and Humidifier from the 
Water Vessel. Take out the Water Vessel.
c) As shown in Figure 8.2.1-d, empty the Vessel. Brush off the mineral substance such as water scale
or verdigris on the Vessel, Water Level Sensors and Humidifier. If the filter of the Humidifier fails to
work, replace with a new filter.
d) Install Water Level Sensor and Humidifier, place the water vessel back and secure the cover plate.

Press "SET" key to remove disengage alarms. See fig. 7.11.1. 

Humidifier
Water Level

Sensor

Highest Level

of Vessel
Water Vessel

Humidifier

Figure 8.2.1-b

Figure 8.2.1-a

8. Disinfec�on and Steriliza�on Control Part

9. Illumina�on Intensity Control

10. O

2 concentra�on monitoring

Press illumination adjustment key "+/-" to adjust 
ten grades of illumination intensity. See fig. 7.9.1. 
The highest is grade nine and the lowest is 
closed. Under the Initial start-up status, it is 
closed.

Key control: press "SET" key for five seconds to enter the parameters setting status. The humidity / O

2

 

display shows flashing digits of O

2

 concentration set value. When the indicator lamp "P" (O

2

 concentra-

tion) is flashing, it is able to adjust the set value of O

2

 concentration alarm. Now press humidity "+/-" 

key to adjust O

2

 set value by ±1 stepping. The default factory setting is the alarm set point of 21%.

a) On: before starting the sterilization function, make sure that there is no pet in the ICU and the ICU

door is closed. Meanwhile, press "SET" key plus "sterilization" key to enter the disinfection mode. See

fig. 7.8.1.

b) Operation: under this mode, all other function keys are void except "SET" key and "sterilization" key. 

Both temperature display and humidity display show "DISI" (abbreviated disinfection); under the

disinfection mode, switch on the UV disinfection lamp. There is violet light inside the ICU (with weak

acrid scent). Meanwhile, start the left fan and the right fan to effect ventilation inside.

c) Off: press "SET" key plus "disinfection" key at the same time again to switch off the UV disinfection

lamp, the disinfection indication lamp and the left fan to exit the disinfection mode and recover the

normal work mode. At this time all function keys are valid except the disinfection key.
Note: when the ICU is under the normal work mode, it is invalid to press "disinfection" key individually.

 Figure 7.9.1

 Figure 7.10.1

 Figure 7.11.1

Hold for 5 seconds

Press “+/-” for se�ng the 

valne for o

2

 alarm

Press then remove the alarm

Set illumina�on

Press “+”, for brighter

Press “-”, for dimmer ten stage adjustable

14

15

Cleaning and Maintenance

Содержание J1550

Страница 1: ...J1550 JorVet ICU User Manual Professional Veterinary Equipment Supplier Jorgensen Laboratories Inc 1450 N Van Buren Ave Loveland CO 80538 800 525 5614 info jorvet com www jorvet com...

Страница 2: ...s moderately sized piece of pet medical equipment will easily fit into most veterinarians practices and pet hospitals Its characteristics and compact structural design function perfectly to serve its...

Страница 3: ...tank has water half of the maximum level 1 Provides bacteria free and thermostatic incubation cocoon for young pets 2 Allows for body temperature recovery fluid infusion atomization treatment incubati...

Страница 4: ...er on the left side of ICU Insert the power plug into the socket on the ICU 2 Screw off the cup lid and nozzle pour medicine liquid into the nebulizer cup 3 Screw on the cup lid and nozzle and put the...

Страница 5: ...5KVd c 3 5KV DC 0 5KV DC Anion concentration At DC12V 6 106 PCS cm3 10 Insulation resistance Power line external box 20M 500V DC Dielectric streangth Power line external box AC 1800V 60HZ 5mA 3S No b...

Страница 6: ...than 2 microns medicines can enter into the finest bronchus and alveoli The JorVet J1550 ICU Unit can produce 0 5 4 micron atomized particles of medicine Meanwhile intelligent control quantity of atom...

Страница 7: ...nd cannot exceed the set value for three minutes It will trigger the alarm system and start the intelligent ventilation system at the same time being ventilated with the outside air and thus improve t...

Страница 8: ...umidifying If an additionally purchased cooler is connected start the cooling function for dehumidifying Indicator lamp control The humidity indicator lamp is always on at the time of humidifying The...

Страница 9: ...on and Steriliza on Control Part 9 Illumina on Intensity Control 10 O2 concentra on monitoring Press illumination adjustment key to adjust ten grades of illumination intensity See fig 7 9 1 The highe...

Страница 10: ...he humidifier For separate cleaning of the humidifier it is not necessary to remove control racks See the chapter of separate cleaning of the humidifier above If the humidifier is definitely damaged o...

Страница 11: ...r Atomizing time and volume depend on concentration of atomizing liquid Repeat the atomization operation procedure if not completely atomized once The humidity is not increased after raising the set p...

Страница 12: ...icitly expressed quality conditions Advance replacement warranty service Your warranty service may involve advance replacement warranty service If the hardware is found to be defective after diagnosis...

Отзывы: