Jorvet J1550 Скачать руководство пользователя страница 3

1.

Please read the Instruction Manual carefully before using the machine. Nurses (and other pet care

workers) should be trained to operate the machine under the guidance of qualified medical care personnel

who are familiar with the functions and understand the risks and advantages of

 

the machine.

2.

 Please count the parts and spare parts when unpacking, and read the Instruction Manual carefully.

3.

The equipment shall be placed in a clean place with limited changes in temperature and humidity, protected 

from direct sunlight and away from heat radiation.

4.

To ensure the proper operation of the Pet ICU, please fill the bottle with distilled water or purified water.

5.

After the machine is turned on, please preset the temperature value, and warm it up for 3-5 minutes before 

use.

  Do not use tap water.

6.

Please do not load weight o

n top of 

the machine, and the maximum load capacity of the machine housing is 

1-2 kilograms (2.2 to 4.45lbs).

7

.

The machine is equipped with an intelligent alarm system sensor for five faults; extreme temperature,

excessive moisture, indoor 

oxygen

 concentration, lack-of-water in the tank and power supply Interrup-tion. 

This alarm system is tested and proven to be safe and reliable. If the machine fails, please inspect the

 

systems immediately to find out reasons and eliminate the faults. If the faults cannot be eliminated in a 

short time, please stop medical treatment or care for pets, and ask for professionals or personnel from the 

JorVet

 Company to repair the machine.

8.   A

llow for the machine to have proper ventilation. Do not block the air outlet and return air inlet.

9.

 After being powered on, if the machine is not used, in order to ensure safety and energy-saving,

 

please

 

turn the switch to off position and allow for sufficient power down before turning the switch back to on 

position.

10.  

Do not use alcohol or other organic solvents to wipe the thermostat cover of the tank. Keep it

 

moisture

 

free. 

11.  

The nebulizer should be unplugged as soon as the nebulization is over.

12.  

When the LAC alarm can be seen on the humidity display panel, the tank inside should be filled to the 

maximum level even if the tank has water half of the maximum level.

1. Provides bacteria free and thermostatic incubation cocoon for young pets.
2. Allows for body temperature recovery, fluid infusion, atomization treatment, incubation, health
improving and comfortable, stable stay in the hospital (for observation for sick, dying and weak pets).

Built-in humidifier bottle: some parts of medical atomizer and the tank are separately packed. Please 
pay attention not to damage or miss any parts or spare parts when unpacking.

JorVet J1550 ICU Unit 

 cabinet

Nebulizer

Power 

cord

 Waste tray 

Water purification filter material (Phosphorus-silicon crystal) User Manual

 

External Humidifier (optional)

As shown in Figure 4.2.1, the Humidifier Vessel is installed in the right control chamber of the 

JorVet 

J1550 ICU Unit

. As shown in Figure 4.2.2, open the outlet cover, take out the built-in Water Vessel, fill 

the built-in Water Vessel with water, (

using 

distilled water or purified water)as shown In Figure 4.2.2. 

And then put back the vessel and place the cover in the proper position.

Notes:

a) The Humidifier and Water Level Sensor should be placed securely onto the vessel to prevent the

spill of water.

b) Do not fill the vessel with purified or distilled water, which may cause the malfunction of the Water

Level Control of the water tank and may eventually cause water to spill or leak.

c) The water level should not be above the Highest Level Line (HLV).

d) The purpose of the water purification filter material is to remove hard water scale and improve the

water quality. The validity period of each placement is 10-12 months. Colorless, tasteless and non-tox-

ic, the filter material meets the National Sanitary Standard for Drinking Water& GB1499- 1988 and can 

effectively improve water quality.

2. Installation of Humidifier Vessel

Outlet Cover

Humidifier
Water Level

Sensor

Highest Level

of Vessel
Water Vessel

Figure 4.2.1 

    Figure 4.2.2 

02

03

Summary of Important Notes

Product Use

Product Installation

The external Nebulizer can be installed on the left 

side of 

JorVet J1550 ICU Unit

. Insert the power plug 

into the socket on ICU; insert the nozzle of 

Nebulizer into the port on the left side as showed in 

Figure 9.1.1.

Before nebulization, switch on the nebulizer by 

pressing the switch on the nebulizer and then turn 

to the operation panel of ICU.

3. Nebulizer Installa�on

    Figure 9.1.1

Содержание J1550

Страница 1: ...J1550 JorVet ICU User Manual Professional Veterinary Equipment Supplier Jorgensen Laboratories Inc 1450 N Van Buren Ave Loveland CO 80538 800 525 5614 info jorvet com www jorvet com...

Страница 2: ...s moderately sized piece of pet medical equipment will easily fit into most veterinarians practices and pet hospitals Its characteristics and compact structural design function perfectly to serve its...

Страница 3: ...tank has water half of the maximum level 1 Provides bacteria free and thermostatic incubation cocoon for young pets 2 Allows for body temperature recovery fluid infusion atomization treatment incubati...

Страница 4: ...er on the left side of ICU Insert the power plug into the socket on the ICU 2 Screw off the cup lid and nozzle pour medicine liquid into the nebulizer cup 3 Screw on the cup lid and nozzle and put the...

Страница 5: ...5KVd c 3 5KV DC 0 5KV DC Anion concentration At DC12V 6 106 PCS cm3 10 Insulation resistance Power line external box 20M 500V DC Dielectric streangth Power line external box AC 1800V 60HZ 5mA 3S No b...

Страница 6: ...than 2 microns medicines can enter into the finest bronchus and alveoli The JorVet J1550 ICU Unit can produce 0 5 4 micron atomized particles of medicine Meanwhile intelligent control quantity of atom...

Страница 7: ...nd cannot exceed the set value for three minutes It will trigger the alarm system and start the intelligent ventilation system at the same time being ventilated with the outside air and thus improve t...

Страница 8: ...umidifying If an additionally purchased cooler is connected start the cooling function for dehumidifying Indicator lamp control The humidity indicator lamp is always on at the time of humidifying The...

Страница 9: ...on and Steriliza on Control Part 9 Illumina on Intensity Control 10 O2 concentra on monitoring Press illumination adjustment key to adjust ten grades of illumination intensity See fig 7 9 1 The highe...

Страница 10: ...he humidifier For separate cleaning of the humidifier it is not necessary to remove control racks See the chapter of separate cleaning of the humidifier above If the humidifier is definitely damaged o...

Страница 11: ...r Atomizing time and volume depend on concentration of atomizing liquid Repeat the atomization operation procedure if not completely atomized once The humidity is not increased after raising the set p...

Страница 12: ...icitly expressed quality conditions Advance replacement warranty service Your warranty service may involve advance replacement warranty service If the hardware is found to be defective after diagnosis...

Отзывы: