Jorvet J1550 Скачать руководство пользователя страница 11

After the power switch is pressed, 

there is no reaction or the power 

indicator lamp is still off

Phenomenon

18 

19

1.Bad power connection and

plug connection

2.Damaged switch

Cause analysis

Solution

1. Check power connection

2.Replace the power switch

Little atomizing volume form 

the medical atomizer

1.Too big atmizing particles of

solution causes little atomizing

volume

2.Solution contains syrup or

colloform

3.Tablets are used for

dissolution

4.Oily impurities on the surface

of solution

1.Too big particles are not

suitable for operation of this

atomizer

2.When solution contains

syrup or colloform,it requires 

with dry filter paper

3.When tablets are used for

dissolution,dissolve sediments

to remove impurites first

4.Remove oily impurities on

the surface of solution with

clean tissue firstivvv

No atomizing from the atomizer

1.Water shortage of the water

tank the atomizer

2.Damaged ultrasonic chip of

the atomizer

3.Water scale in the tank of the

atomizer

1.Add water into the water

tank until it is at the same

level of the cup bottom

2.Replace the ultrasonic chip

3.Gently scrub the inside of

the tank with lint

Remaining solution not 

atomized in the cup after 

atomization of the medical 

atomizer

Atomizing time and volume 

depend on concentration of 

atomizing liquid

Repeat the atomization 

operation procedure if not 

completely atomized once

The humidity is not increased 

after raising the set point of 

humidity

1.Water shortage in the bottle

of the humidifier

2.Damaged chip of the

humidifier

3.Water scale in the metal

water level sensor inside the

humidifier tank

1.Add water into the bottle

2.Replace the humidifier

3.Remove water scale inside

according to the chapter of

“separate cleaning of the

humidifier”, or try to dip some

vinegar into the water bottle of

the humidifier

Operational keys on the 

panel do not work

Bad contact of panel 

plugins

Damaged panel keys

1.Install plug-ins properly

2.Replace the panel

No heating

1.Failed or damaged fan

2.Heater (heat sink) with

bad contact or damaged

3.ICU door opened

1.Check and replace fan

2.Replace heater band or

heat sink

3.Close ICU door

No cooling

1.No external cooler

connected

2.Cooling pipe of the cooler

not connected to the tank

3.Signal line of the cooler

not connected to the

machine

1.Connect it to an externally

purchased cooler

2.Check the cooling pipe

3.Plug the signal line into the

colling socket of the tank

Continuous alarm for O

2

 

concentration

1.Failed or damaged fan

2.Not enough instant ventila-

tion inside the chamber

3.damaged O

2

 sensor

1.Chech fans

2.Open the ICU door

3.Replace the O

2

 concentra-

tion sensor

Troubleshooting  

Содержание J1550

Страница 1: ...J1550 JorVet ICU User Manual Professional Veterinary Equipment Supplier Jorgensen Laboratories Inc 1450 N Van Buren Ave Loveland CO 80538 800 525 5614 info jorvet com www jorvet com...

Страница 2: ...s moderately sized piece of pet medical equipment will easily fit into most veterinarians practices and pet hospitals Its characteristics and compact structural design function perfectly to serve its...

Страница 3: ...tank has water half of the maximum level 1 Provides bacteria free and thermostatic incubation cocoon for young pets 2 Allows for body temperature recovery fluid infusion atomization treatment incubati...

Страница 4: ...er on the left side of ICU Insert the power plug into the socket on the ICU 2 Screw off the cup lid and nozzle pour medicine liquid into the nebulizer cup 3 Screw on the cup lid and nozzle and put the...

Страница 5: ...5KVd c 3 5KV DC 0 5KV DC Anion concentration At DC12V 6 106 PCS cm3 10 Insulation resistance Power line external box 20M 500V DC Dielectric streangth Power line external box AC 1800V 60HZ 5mA 3S No b...

Страница 6: ...than 2 microns medicines can enter into the finest bronchus and alveoli The JorVet J1550 ICU Unit can produce 0 5 4 micron atomized particles of medicine Meanwhile intelligent control quantity of atom...

Страница 7: ...nd cannot exceed the set value for three minutes It will trigger the alarm system and start the intelligent ventilation system at the same time being ventilated with the outside air and thus improve t...

Страница 8: ...umidifying If an additionally purchased cooler is connected start the cooling function for dehumidifying Indicator lamp control The humidity indicator lamp is always on at the time of humidifying The...

Страница 9: ...on and Steriliza on Control Part 9 Illumina on Intensity Control 10 O2 concentra on monitoring Press illumination adjustment key to adjust ten grades of illumination intensity See fig 7 9 1 The highe...

Страница 10: ...he humidifier For separate cleaning of the humidifier it is not necessary to remove control racks See the chapter of separate cleaning of the humidifier above If the humidifier is definitely damaged o...

Страница 11: ...r Atomizing time and volume depend on concentration of atomizing liquid Repeat the atomization operation procedure if not completely atomized once The humidity is not increased after raising the set p...

Страница 12: ...icitly expressed quality conditions Advance replacement warranty service Your warranty service may involve advance replacement warranty service If the hardware is found to be defective after diagnosis...

Отзывы: