background image

Joerns

®

 EasyCare

®

 Bed Platform

Model ECS

14

14

To Remove Actuator

Recommended Tools: Screwdriver, Hammer,  

Masking Tape, Zip Ties

1.  Identify the actuator to be replaced.

2.  Tip the bed on its side to replace high/low 

actuator(s).

3.  Unplug the actuator cord.

4.  The actuator is held in place by two pins. Use 

a screwdriver to remove the lock ring on each 

pin.

5.  Slide the pins out of holes. If necessary, use 

the hammer to tap out the pins.

6.  You may now replace the actuator(s).

7.  To reassemble the bed, reverse the previous 

steps, making note of the following two 

important items:

•  Assemble the pin and lock rings as  

originally installed.

•  Zip-ties should be replaced, with cords 

returned to their original position and 

routing direction to the control box.

To Remove AC Power Cable

Recommended Tools: Small Screwdriver,  

Masking Tape, Zip Ties, 5/16" Nut Driver

1.  Raise the bed so that there is about 27 inches 

of clearance between the floor and the mattress 

support platform of bed.

2.  Unplug the power cable from the outlet.

3.  Cut the zip-ties that holds the power cable to 

the frame. Tape the remaining cables in their 

original position. It is very important to keep the 

actuator cables routed along the frame in their 

original factory-installed positions. If this step is 

not followed, damage to cables or electric shock 

may result.

4.  Detach the ground wire from the bed frame  

if equipped. 

 

Warning: Possible Shock Hazard.

 

Injury may result from improper power cable 

routing. Always follow proper factory-installed 

routing configuration. Be sure to check condition 

of the power cable on a regular basis. Do not 

use damaged or frayed power cables. 

5.  Using a small screwdriver, alternately depress 

the small red latch tabs above and below AC 

receptacle on the control box. Simultaneously, 

pull gently outward on the power cable where it 

connects to the control box, until the plug comes 

out.

6.  Plug the replacement cable into the control box 

until it latches. Attach the ground wire to the bed 

frame.

7.  Re-route the power cable along the frame in the 

same position as the original routing.

8.  Replace zip-ties along with cables in their 

original position and routing to the control box. 

If this step is not followed, damage to the cables 

or electric shock may result.

Содержание ECS EasyCare 3

Страница 1: ...Joerns Healthcare EasyCare 5 EasyCare 7 EasyCare 9 User Service Manual Joerns EasyCare Bed Platform Model ECS leading bed evolution To avoid injury read user s manual before using EasyCare 3...

Страница 2: ...ble Shock Hazard Injury may result from improper routing of the power cord Always follow the proper factory installed routing configuration Failure to do so could result in injury or death Warning Pos...

Страница 3: ...less than 1 5 inches under user s weight Warning Possible Injury or Death Floor locks increase bed stability and resident safety Make sure the bed is in a locked position with respect to the floor by...

Страница 4: ...althcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance Solutions Matching the right bed components in order to meet regulatory guidelines...

Страница 5: ...Sensor EasyCare 9 Power 120 VAC Frequency 60 Hz Current Rating 4 0 Amps 120V Classification Class I Type B or Class II Type B Mode of Operation 10 Maximum Duty Cycle 2 minutes on 18 minutes off Circui...

Страница 6: ...face down 3 The foot panel brackets are the shorter of two sizes Figure 1 4 Install the foot panel brackets with the four screws provided The bracket offset should face up Figure 2 Do not tighten unti...

Страница 7: ...e mattress side stop into the hole on one side 6 Compress the mattress side stop and insert into the opposite hole on the head end of the mattress support platform IV Sockets Joerns EasyCare Bed Platf...

Страница 8: ...15 Operation Pendant Operation EasyCare 3 5 and 7 Your Joerns EasyCare 3 EasyCare 5 or EasyCare 7 bed comes standard with the snap in pendant The pendant has raised high contrast buttons and graphics...

Страница 9: ...e 19 Figure 20 Figure 21 Figure 22 EasyCare 9 1 Your EasyCare 9 bed comes standard with a pendant Figure 18 The pendant plugs into a port under either side of the seat pan 2 Unscrew the cap Figure 19...

Страница 10: ...ock graphic on the panel and push the function to unlock Amber LED is no longer illuminated signaling that the function is now operable Pendant Lockout and Bed Operation 1 Push button to lock out Back...

Страница 11: ...cked position by a red symbol Figure 28 3 To unlock the UltraLock place the ball of your foot on any pedal and push down Figure 29 The label indicates the mobile position by a green symbol Figure 28 W...

Страница 12: ...bed function fails to operate Ensure that the bed is not overloaded or constrained from movement Actuator and Control Box Information Cord and Socket Identification Figures 31 and 32 1 Back Actuator...

Страница 13: ...Control Box contain no serviceable parts Overload protection is achieved via current sensing technology which will automatically reset If any bed function fails to operate check to see if the bed is o...

Страница 14: ...ties that holds the power cable to the frame Tape the remaining cables in their original position It is very important to keep the actuator cables routed along the frame in their original factory ins...

Страница 15: ...es or less Do not surpass 10 duty cycle 2 min on 18 min off 2 If these steps fail to resolve the problem call Technical Support at 800 826 0270 1 Switch the pendant with a new pendant or one from a be...

Страница 16: ...ght in these locations Failure to do so could increase risk of suffocation or result in injury or death Warning Possible Injury The extender is designed for use with a foot panel Sitting on or sliding...

Страница 17: ...itable for use with the EasyCare Bed due to its low operating height 5 Adjustment and use of a Trapeze should be in accordance with instructions from a healthcare professional doctor or therapist 6 Do...

Страница 18: ...rest on the floor particularly when the bed is in its lowest position Otherwise a safety condition may occur or the function of attached items may be impacted Bed Transport Bar Figure 40 1 The Bed Tra...

Страница 19: ...he Battery will be fully charged Note We recommend that users charge up the Battery immediately upon receipt 4 The Battery has a useful life of about 3 5 years depending upon usage It is recommended t...

Страница 20: ...s Joerns beds contain various parts that wear from normal use These parts such as DC batteries are not covered under the three year warranty but are covered for 90 days after date of delivery Joerns H...

Отзывы: