background image

Joerns

®

 EasyCare

®

 Bed Platform

Model ECS

10

Figure 25

Staff Control Panel Operation

 

(Optional)

EasyCare

®

 3, 5 and 7 (Figure 23)

The Staff Control Panel allows caregivers to lock 

out all of the bed functions individually. Caregivers 

can also operate the bed from this panel without 

walking to side of the bed. 

1.  To lock out a function, press panel’s “no 

pendant” (Figure 23: #1, #3 and #5) icon. 

Illuminated red LED light indicates that the 

function is now locked out. 

2.  To unlock a function, press panel’s “pendant” 

(Figure 23: #2, #4 and #6) icon. Red LED is no 

longer illuminated, signaling that the pendant 

can now operate that specific function.

EasyCare 9 (Figure 24)

The Staff Control Panel allows the caregivers 

to operate the bed from the foot panel without 

walking to the bedside. The caregiver can also 

individually lock out all functions, both Pendant 

and Staff Control. 

1.  To lock out a function, press and hold the 

“Lock” graphic on the panel and push the 

function to lock out. Illuminated amber LED 

indicates that the function is now locked out. 

2.  To unlock a function, press and hold the “Lock” 

graphic on the panel and push the function to 

unlock. Amber LED is no longer illuminated, 

signaling that the function is now operable.

Pendant Lockout and Bed Operation

1.  Push button to lock out Back

2.  Push button to unlock Back

3.  Push button to lock out Knee

4.  Push button to unlock Knee

5.  Push button to lock out Hi/Lo

6.  Push button to unlock Hi/Lo

7.  Push button to raise Back

8.  Push button to lower Back

9.  Push button to raise Knee

10. Push button to lower Knee

11.  Push button to raise Bed

12. Push button to lower Bed

2

4

6

5

12

11

1

3

7

8

9 10

Figure 23

Standard Staff Control

1.  Push button to raise Back

2.  Push button to lower Back

3.  Push button to raise Knee

4.  Push button to lower Knee

5.  Push button to raise Bed

6.  Push button to lower Bed

7.  Push and hold to change lockout status in 

combination with another function button.

2

1

4

6

5

3

Figure 24

7

Roll in Low Lock System

EasyCare 3 and 5

1.  To keep the bed from rolling when in the lowest 

position, press down on the locking lever to lock 

casters (Figure 25).

2.  The standard caster set comes with two locking 

casters placed at opposing corners of bed.

Mobile 

Position

Locked  

Position

Swivel 

Casters

Roll in 

Low Lock 

System

Low only

Any Height 

(casters must 

be locked in 

Low)

Head & 

Foot (in Low 

position 

only)

Содержание ECS EasyCare 3

Страница 1: ...Joerns Healthcare EasyCare 5 EasyCare 7 EasyCare 9 User Service Manual Joerns EasyCare Bed Platform Model ECS leading bed evolution To avoid injury read user s manual before using EasyCare 3...

Страница 2: ...ble Shock Hazard Injury may result from improper routing of the power cord Always follow the proper factory installed routing configuration Failure to do so could result in injury or death Warning Pos...

Страница 3: ...less than 1 5 inches under user s weight Warning Possible Injury or Death Floor locks increase bed stability and resident safety Make sure the bed is in a locked position with respect to the floor by...

Страница 4: ...althcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance Solutions Matching the right bed components in order to meet regulatory guidelines...

Страница 5: ...Sensor EasyCare 9 Power 120 VAC Frequency 60 Hz Current Rating 4 0 Amps 120V Classification Class I Type B or Class II Type B Mode of Operation 10 Maximum Duty Cycle 2 minutes on 18 minutes off Circui...

Страница 6: ...face down 3 The foot panel brackets are the shorter of two sizes Figure 1 4 Install the foot panel brackets with the four screws provided The bracket offset should face up Figure 2 Do not tighten unti...

Страница 7: ...e mattress side stop into the hole on one side 6 Compress the mattress side stop and insert into the opposite hole on the head end of the mattress support platform IV Sockets Joerns EasyCare Bed Platf...

Страница 8: ...15 Operation Pendant Operation EasyCare 3 5 and 7 Your Joerns EasyCare 3 EasyCare 5 or EasyCare 7 bed comes standard with the snap in pendant The pendant has raised high contrast buttons and graphics...

Страница 9: ...e 19 Figure 20 Figure 21 Figure 22 EasyCare 9 1 Your EasyCare 9 bed comes standard with a pendant Figure 18 The pendant plugs into a port under either side of the seat pan 2 Unscrew the cap Figure 19...

Страница 10: ...ock graphic on the panel and push the function to unlock Amber LED is no longer illuminated signaling that the function is now operable Pendant Lockout and Bed Operation 1 Push button to lock out Back...

Страница 11: ...cked position by a red symbol Figure 28 3 To unlock the UltraLock place the ball of your foot on any pedal and push down Figure 29 The label indicates the mobile position by a green symbol Figure 28 W...

Страница 12: ...bed function fails to operate Ensure that the bed is not overloaded or constrained from movement Actuator and Control Box Information Cord and Socket Identification Figures 31 and 32 1 Back Actuator...

Страница 13: ...Control Box contain no serviceable parts Overload protection is achieved via current sensing technology which will automatically reset If any bed function fails to operate check to see if the bed is o...

Страница 14: ...ties that holds the power cable to the frame Tape the remaining cables in their original position It is very important to keep the actuator cables routed along the frame in their original factory ins...

Страница 15: ...es or less Do not surpass 10 duty cycle 2 min on 18 min off 2 If these steps fail to resolve the problem call Technical Support at 800 826 0270 1 Switch the pendant with a new pendant or one from a be...

Страница 16: ...ght in these locations Failure to do so could increase risk of suffocation or result in injury or death Warning Possible Injury The extender is designed for use with a foot panel Sitting on or sliding...

Страница 17: ...itable for use with the EasyCare Bed due to its low operating height 5 Adjustment and use of a Trapeze should be in accordance with instructions from a healthcare professional doctor or therapist 6 Do...

Страница 18: ...rest on the floor particularly when the bed is in its lowest position Otherwise a safety condition may occur or the function of attached items may be impacted Bed Transport Bar Figure 40 1 The Bed Tra...

Страница 19: ...he Battery will be fully charged Note We recommend that users charge up the Battery immediately upon receipt 4 The Battery has a useful life of about 3 5 years depending upon usage It is recommended t...

Страница 20: ...s Joerns beds contain various parts that wear from normal use These parts such as DC batteries are not covered under the three year warranty but are covered for 90 days after date of delivery Joerns H...