5. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The ElectroVita is intended for use in the electromagnetic environment specified below. The 
customer or the user of the ElectroVita should assure that it is used in such an environment.

Immunity Test

IEC 60601

Compliance

Electromagnetic Environment Guidance

Test Level

Level

Electrostatic

±6 kV contact

±6 kV contact

Floors should be wood, concrete or ceramic tile.

discharge (ESD)

±8 kV air

±8 kV air

If oors are covered with synthetic material, the

IEC 61000-4-2

relative humidity should be at least 30 %.

Electrical fast

±2 kV for power

±2 kV for power

Mains power quality should be that of a typical

transient/burst

supply lines

supply lines

commercial or hospital environment.

IEC 61000-4-4

±1kV for

±1kV for

input/output

input/output

lines

lines

Surge

±1 kV line(s)

±1 kV line(s)

Mains power quality should be that of a typical

IEC 61000-4-5

and neutral

and neutral

commercial or hospital environment.

Voltage dips, short

<5 % UT

<5 % UT

Mains power quality should be that of a typical

interruptions and

(>95 % dip in UT)

(>95 % dip in UT) commercial or hospital environment. If a dips or an

voltage variations

for 0,5 cycle

for 0,5 cycle

interruption of mains power occurs, the current of the

on power supply

40 % UT

40 % UT

ElectroFlex may be dropped o from normal level, it

input lines

(60 % dip in UT)

(60 % dip in UT)

may be necessary to use uninterruptable power

IEC 61000-4-11

for 5 cycles

for 5 cycles

supply or a battery

70 % UT

70 % UT

(30 % dip in UT)

(30 % dip in UT)

for 25 cycles

for 25 cycles

<5 % UT

<5 % UT

(>95 % dip in UT)

(>95 % dip in UT)

for 5s

for 5s

Power frequency

3 A/m

Not applicable

Not applicable

(50/60 Hz)
magnetic eld
IEC 61000-4-8

NOTE: UT is the AC mains voltage prior to application of the test level

6. Guidance and manufacturer’s declaration – electromagnetic immunity

The ElectroVita is intended for use in the electromagnetic environment specified below. The 
customer or the user of the ElectroVita should assure that it is used in such an environment.

19

Содержание circulation enhancer

Страница 1: ...Instructions...

Страница 2: ...ired nerves muscles and capillaries Capillaries are the smallest of your bodies blood vessels and provide the exchange between cells and circulation So relax and re energise with the ElectroVita Circu...

Страница 3: ...display will light up blue and remain on 3 Remove your footwear and socks or stockings Your feet need to be completely bare to experience the ElectroVita micro current stimulation massage therapy 4 Th...

Страница 4: ...hing that is compulsory must be observed Compulsory actions are marked clearly with this warning symbol This product should not be used by persons with electronic medical implants e g heart pacemakers...

Страница 5: ...with ointments including spray type ointments May cause discomfort or ill health Simultaneous connection of a PATIENT to h f surgical EQUIPMENT may result in burns at the site of the STIMULATOR electr...

Страница 6: ...n under treatment may become exhausted and may cause poor physical condition Put the Long Life pads only on skin or on the storage tray to avoid damage of the adhesive surfaces of the pads Correct Dis...

Страница 7: ...both immunity and emissions Nevertheless special precautions need to be observed Do not use mobile cellular telephones and other devices which generate strong electrical or electromagnetic elds near...

Страница 8: ...mulator is intended to be used as a massager to relieve muscle pain increase blood circulation relax stiff muscles reduce swollen feet ankles and relieve fatigue The massaging effect is achieved by el...

Страница 9: ...pare Built in Ergonomic Carrying Handle Plastic Protector for Gel Pads Electrode Gel Pads to target speci c areas of pain and swelling Output for Body for connecting Gel Pads cable jack for Body Massa...

Страница 10: ...one of the 10 pre programmed massage programs Downward D Increase the output intensity of foot BAND 1 FROM 1 40 LEVEL E Decrease the output intensity of foot BAND 1 FROM 1 40 LEVEL F Increase the out...

Страница 11: ...repeating 2 Pulse rate 12Hz 80Hz with 12 seconds and o time in 1 sec A cycle repeating 3 Pulse rate 7Hz with 30 seconds and o time in 1 sec A cycle repeating 4 Pulse rate 7Hz 25Hz with 8 seconds and...

Страница 12: ...ElectroVita output wave forms 11...

Страница 13: ...In order to let the user feel the stimulation gradually increase the intensity to a comfortable level by holding down the intensity button Those who suffer with poor circulation may not feel anything...

Страница 14: ...ee Figure 8 The LCD will show the level selected see Figure 9 When you reach the desired Level the icon will blink To stop the blinking and maintain the intensity press the button 5 times 6 To termina...

Страница 15: ...y aligned Note on batteries Do not mix different types of batteries or an old battery with a new one To prevent the risk of leakage or explosions never recharge the batteries apply heat or take them a...

Страница 16: ...JDF DMFBO CZ VTJOH B TPGU DMPUI UP DMFBO UIF TVSGBDF PG UIF VOJU t 5P DMFBO UIF GPPU QMBUFT VTF B TPGU EBNQ DMPUI ESZ UIPSPVHIMZ BGUFS DMFBOJOH t 5P DMFBO UIF NBJO VOJU VTF B TPGU EBNQ DMPUI ESZ UIPSP...

Страница 17: ...d Adhesive surface Use of gel pad during perspiring Chill the gel pad in a domestic fridge for 3 4 hours of gel pad is sticky Gel pad washed too long and or too frequently Gel pads stored under high t...

Страница 18: ...100mA 0 45VA Battery Life Approx 1 Month When used 25 minutes a day continuously Frequency Generation Approx 1Hz to 80Hz Power Consumption 0 45W Maximum Output Voltage U 50V during 1k load Maximum Ou...

Страница 19: ...equipment Guidance and manufacturer s declaration electromagnetic emissions The ElectroVita is intended for use in the electromagnetic environment speci ed below The customer or the user of the Elect...

Страница 20: ...IEC 61000 4 5 and neutral and neutral commercial or hospital environment Voltage dips short 5 UT 5 UT Mains power quality should be that of a typical interruptions and 95 dip in UT 95 dip in UT comme...

Страница 21: ...ach frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not...

Страница 22: ...ording to frequency of transmitter output power m of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz W d 1 2 P d 1 2 P d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 1...

Страница 23: ...22...

Страница 24: ...JML Distributed by JML Unit 1 Eastside Port of Tyne South Shields NE33 5SP info JMLgroup co uk Freephone 0800 781 7831 Web www JMLdirect com Registered address JML House NW5 3EG Product code V0393...

Отзывы: