5. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ElectroVita is intended for use in the electromagnetic environment specified below. The
customer or the user of the ElectroVita should assure that it is used in such an environment.
Immunity Test
IEC 60601
Compliance
Electromagnetic Environment Guidance
Test Level
Level
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood, concrete or ceramic tile.
discharge (ESD)
±8 kV air
±8 kV air
If oors are covered with synthetic material, the
IEC 61000-4-2
relative humidity should be at least 30 %.
Electrical fast
±2 kV for power
±2 kV for power
Mains power quality should be that of a typical
transient/burst
supply lines
supply lines
commercial or hospital environment.
IEC 61000-4-4
±1kV for
±1kV for
input/output
input/output
lines
lines
Surge
±1 kV line(s)
±1 kV line(s)
Mains power quality should be that of a typical
IEC 61000-4-5
and neutral
and neutral
commercial or hospital environment.
Voltage dips, short
<5 % UT
<5 % UT
Mains power quality should be that of a typical
interruptions and
(>95 % dip in UT)
(>95 % dip in UT) commercial or hospital environment. If a dips or an
voltage variations
for 0,5 cycle
for 0,5 cycle
interruption of mains power occurs, the current of the
on power supply
40 % UT
40 % UT
ElectroFlex may be dropped o from normal level, it
input lines
(60 % dip in UT)
(60 % dip in UT)
may be necessary to use uninterruptable power
IEC 61000-4-11
for 5 cycles
for 5 cycles
supply or a battery
70 % UT
70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles
for 25 cycles
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5s
for 5s
Power frequency
3 A/m
Not applicable
Not applicable
(50/60 Hz)
magnetic eld
IEC 61000-4-8
NOTE: UT is the AC mains voltage prior to application of the test level
6. Guidance and manufacturer’s declaration – electromagnetic immunity
The ElectroVita is intended for use in the electromagnetic environment specified below. The
customer or the user of the ElectroVita should assure that it is used in such an environment.