PAGE 14 of 35
3.
Loosen the probe holder adjustment knob and remove the outer probe
holder arm
(Fig. 36)
.
4.
Loosen the arm clamp screw
(Fig. 37).
5.
Place the wedge on the pivot button of the inner probe holder arm
(Fig. 37).
6.
Align the middle of the wedge with the centre of the yoke
(Fig. 38).
7.
Tighten both the probe holder adjustment knob and the arm clamp screw
(Fig. 39)
while ensuring the wedge remains centred with the yoke.
4.6.2. Probe Holder Vertical Adjustment
1.
Gently lift the heavy duty probe holder and simultaneously pull the latch
(Fig. 40)
. This action will unlock the probe holder. Slowly lower the probe
holder towards the scan surface
(Fig. 41)
.
Fig. 38
- Remove outer arm
Fig. 39
- Pivot buttons
Fig. 40
- Press up and pull latch
Fig. 41
- Lowered toward scan surface