Measurement accuracy depends on measurement stability, temperature coefficient,
linearity,
repeatability and correction interpolation error.
Checking the measurement accuracy should be taken under the following circumstances:
a
.
Warm
‐
up time should be more than 30 minutes.
b
.
Correctly short the test cables, turn 0 ADJ to ON and perform short calibration by
pressing the touch key or 0 ADJ panel.
The correct short of the test cable is as follows:
4.3.1 Basic Accuracy for Resistance Measurement
JK2516
RANGE
20mΩ
200mΩ
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
Current
1A
1A
100mA
10m
A
1mA
100μA
100μA
10μA
1μA
open
voltage
0.7V
3V
resolution
1μΩ
10μΩ
100μΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
Accuracy
0.1%+
3
0.05%+
2
0.05%+2
0.2%+2
temperatur
e
coefficient
330ppm
100ppm
JK2516A
Range
200m
Ω
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
Current
1A
100mA
10mA
1mA
100μA
100μA
10μA
open
0.7V
3V