12/17
Compare modes:
1) ABS (high and low limits)
In this mode, you can set low and high limits. The instrument will take comparison
between the
tested values and the set ones so as to judge the DUT is HI or LO or IN.
2) % (percent error)
In this mode, you can set the nominal value and the percent error. If you set the nominal
value as
100 and percent as 10, the set value should be 100±10%. The instrument will take
comparison
between the tested values and the set ones so as to judge the DUT is HI or LO or IN.
3.2.3 up super and low super settings
1)Touch the up super, the following soft keys will be displayed.
Press the up and down key to select the value and press the confirmation key to select the
highest setting of 6 digits.
If you find an error in setting, you can select "Backsapce", delete the data by pressing the
key, and select "clear".
Press the confirmation key to reset. After setting up, select "Enter" and click confirm key to
confirm the setting result.
Select "Esc", click the confirmation button to exit the upper limit setting interface, and
enter the measurement display interface.
The setting of the low super is the same as the method of setting the up super.
4)Performance Index
4.1Measurement Function
4.1.1 Measurement Parameters and Notations
R
:
Resistance
LPR: Resistance tested at low voltage
4.1.2 Measurement Groups
Two measurement groups are available : R ,LPR