![Jinko JK2516B Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/jinko/jk2516b/jk2516b_user-manua_2018510014.webp)
13/17
4.1.3 Range
Resistance Mode: AUTO, MANU (HOLD, UP, DOWN)
4.1.4 Trigger
Internal, manual, external, bus
Internal: Continuously test a DUT and then output and display the result.
Manual: Press the “TRIGGER” button on the panel, the instrument will test a DUT once
and display the result. This mode keeps in waiting mode when it is not used.
External: Test a DUT once and display the result when the instrument receives
an external “start up” signal from the footswitch of HANDLER interface on the rear panel.
BUS: The measurement of the instrument will be triggered through the communication
interface.
4.1.5 Mode of Test Terminal
4
‐
terminal measurement mode
DRIVE
HI
:
Current
‐
drive high terminal
DRIVE LO: Current
‐
drive low terminal
SENSE
HI
:
Voltage
‐
sense high terminal
SENSE LO
:
Voltage
‐
sense low terminal
4.1.6 Time Expenditure of Measurement
Measurement time=100ms +t1 /100ms +t1 ( 60Hz ) fast speed
130ms +t1 /130ms +t1 ( 60Hz ) med speed
170ms +t1 /170ms +t1 ( 60Hz ) slow speed
Note: t is the waiting time of measurement, and T1 is the data processing time,
about 5ms.
4.2Test Signal
4.2.1 Current range
JK2516 :Current range: 1μA
‐
1A
JK2516A :Current range:10μA -1A
JK2516B:Current range:100μA-1A
4.2.2 Output Voltage of Open Circuit
Output voltage of open circuit: 0.7V ,3V ,40mV
4.2.3 Maximum Display Range
Parameter
Measurement Display Range
R
1μΩ – 2MΩ
LPR
0
.1mΩ-2.1kΩ
4.3Measurement Accuracy