3
Figure 1-3 Interface
is shown in Figure
1.2 Probe
1 Safety of probe
Set around the probe body protection to protect your fingers to prevent electric shock.
Before taking any measurements, make sure the probe is connected to the oscilloscope
and the ground terminal to ground.( Note: The oscilloscope probe and the attenuation
must be set the same gear)
2 Probe compensation (see probe manual)
In the first probe with any input channel connection, this needs to be adjusted to match the
probe to the input channel. Uncompensated probe calibration can cause measurement
error or errors. If you adjust the probe compensation, as follows:
(1) Set the Probe option attenuation to 10X in the channel menu, The switch on the probe
is also set to 10X,And oscilloscope probe connected to channel 1.If using the probe
hook-tip, ensure reliable contact with the probe.
(2)Connect the probe tip and the signal generator output connector, Grounding clip and
signal generator is connected to the ground connector, Display Channel, Then press the
"AUTO" (automatic) button.
(3)Check the shape of the waveform display. As Figure 1-4
Trigger
status
display
Channel 1 mark
Show vertical
volts / division
Display power
Horizontal
displacement
Show the amount of
the trigger position
Channel
operation
status
Show
horizontal
time / div
Display
level
Trigger
position
Corresponds to a different
function keys, menus will
vary
Frequency
waveform display