SX8 User Manual
G-‐UM-‐00004 Rev 006
38 of 45
[ROUTe]:CLOSe[m]?
Syntax
:[ROUTe]:CLOSe[m]
?<space>
[MAX|MIN]
Function Returns the current channel number on module ‘m’ (‘m’
becomes the new current module number). If m is omitted,
returns the current channel number on the current module.
The following parameters are accepted by the command:
•
no parameter:
returns the current channel number
•
MAX:
returns the maximum configured channel
number
•
MIN:
returns the first channel number
Example
CLOSE?
returns 10.
CLOSE2?
returns 5. (on module 2).
:ROUT:CLOSe2? MAX
returns the maximum configured
channels on module 2.
LCL
Syntax
:LCL
Function Returns the control status to local front panel control. This
function is identical to pressing the “LOCAL” key on the front
panel when the instrument is in the remote control mode.
Example
LCL