BA-40400-02-V14  |  17

Operating instructions

4.10 Operation and keyboard layout

171 mm

267 mm

302 mm

240 mm

a
b
c

d

f

Fig.11: Housing of the TOPAX with operating panel.

Keys

Functions

Arrow key cross

       

•  Menu change in the "Service" menu
•  Change between individual numbers
•  Change numerical values, adjust parameters
Press the key to change from one menu item to 
the other. Numeric values are entered and modi-
fied continuously.

"OK" key between the 
arrow keys

Accept and store inputs
Saving is confirmed by a long beep noise.

Key 

a

: „ESC“

Exit the menu, one level back
Terminate input without saving

Key 

a

: „ESC“

(press down for 5 
seconds)

Return to the main menu

Keys 

b

 - 

f

The functions of the keys vary according to the 
menu and are displayed respectively.

If the device is not operated for approx. five minutes, the controller re-
turns automatically to measuring mode. Changed parameters not con-
firmed with the “OK” key are not saved.
Exception: this time restriction does not apply during calibration and 
configuration.

NOTICE!

If no values or settings are to be changed, 
press "ESC" to exit the selected menus at any 
time. The controller uses the inital settings. 
Values and settings are confirmed 
and saved with the "OK" button. 
The exception to this rule are the "clock" and "timer" 
functions. They can be modified without confirma-
tion.

General input rules:
•  All values can only be changed within the permissible range.
•  Some functions can be protected by a password against unauthor-

ized access (see “12.1.23 Menus 5.9 and 5.10: Code und service 
password” on page 35
).

4.11 First set-up and programming guidelines

After correct installation, switch on the power supply.
The TOPAX is programmed by the manufacturer based on customers’ 
specs before shipment and is supplied along with a terminal connection 
diagram.

IMPORTANT!

All menus of 

TOPAX

 are shown a ID-number marked 

in the bottom line on the left display corner.

4.11.1 Backlight

The backlight of the chloriDES controller is dimmed automatically after 
10 minutes and a blue LED under the arrow key cross is lit.
Once any key is pressed, the LED goes out and the backlight turns on 
again.

4.11.2 Choice of Language

After a short hardware test the during the initial start-up, the TOPAX 
displays the home page for configuration. This screen is shown until a 
language has been chosen by pressing the key 

b

 (German), 

c

 (Eng-

lish), 

 (French) or 

f

 (further languages “>>”).

Deutsch

English

Francais

0.1

Software Version:  ...........    

Fig.12: Home page for the configuration of language choice

4.11.3 Further languages - Memory card

Besides the fixed installed languages of German and English there 
is other language data for the TOPAX. These may be loaded from a 
memory card.
The memory card with the language data is inserted on the display 
board in the upper casing.

4.11.4 Help-Menu

The Help Menu is available as a point of help for many menu points. By 
pressing a random key the TOPAX returns to the previous menu again.

4.11.5 Menu 0.4.1: Input Disinfection

The disinfection connection will be pre-configured according to the or-
der and cannot be changed. 
The following options are possible:
•  2-electrodes measuring cell CS 120
•  Potentiostat (PM)
•  Encapsulated electrode
•  Redox single-rod measuring cell
•  Total chlorine measuring cell
The selected option is highlighted in grey.

Содержание TOPAX DE

Страница 1: ...TOPAX DE Two Channel Controller EN 01 Operating instructions Read this operating manual before using the equipment To be retained for future reference...

Страница 2: ...tion amperometric measurement cells 20 5 2 pH value measurement input 21 5 3 Measurement output Redox potential 23 5 4 Temperature measurement input 23 6 Explanation of digital signal inputs 24 6 1 St...

Страница 3: ...behalf of the operating company The operating company must also ensure that the operating manual has been under stood 1 4 Electrical device safety instructions Basic safety precautions should always b...

Страница 4: ...nd obtaining spare parts TOPAX may be converted or changed only by qualified technical person nel If the configuration of the TOPAX is wrongly made by assembly or ser vice personnel errors and dangers...

Страница 5: ...als Total chlorine measuring input Encapsulated electrode Measuring range adjustable from 0 1 00 mg l 0 2 00 mg l or 0 10 mg l Disturbance variable input option ally available 0 20 mA or 4 20 mA Distu...

Страница 6: ...mm 8 7 mm Relay output alarm M16 X 1 5 NYM O 3x 1 5 mm 9 1 mm connection to PC computer cable Cat 5 M12 X 1 5 Type 2X2XAWG24 1 5 7 mm Connection of current outputs remote communication cable M12 X 1 5...

Страница 7: ...ts are removed at the same time the front housing will no longer be attached to the bottom section ATTENTION The equipment must only be opened when it is dis connected from the power supply 4 2 Dimens...

Страница 8: ...he software configuration automatically assigns these outputs to the respective terminals in a sequential order depending on the configura tion of the TOPAX 1 Disinfection controller 2 Controller pH v...

Страница 9: ...ate the RS 485 resistances on the last TOPAX in the network 4 4 6 TopView software For remote viewing on a PC the manufacturer offers the full version of the TopView display program and a freeware ver...

Страница 10: ...must be clean and dry The permissible bending radius of the cable must be respected The quality of the holding wires must correspond to the defaults from the data sheet of the sensor IMPORTANT If pos...

Страница 11: ...surement output 0 4 20 mA Redox 6 7 Measurement output 0 4 20 mA Temperature or programmed as controller output 8 9 Measurement output 0 4 20 mA combined chlorine or programmed as controller output 10...

Страница 12: ...r polarity of clips 59 and 61 at wish 60 Driver 61 4 7 3 Input module 3x Terminal Function Cable colour 71 Reference electrode with integrated cable Disinfection potentiostatic measuring cell Referenc...

Страница 13: ...71 72 73 7 pH Single Rod Measuring Cell PE110 pH 52 53 8 Redox Single Rod Measuring Cell ME110 ORP 54 55 9 Reference Electrode for Potentiostat SS Electrode for Potentiostat 1 4571 Gold Electrode for...

Страница 14: ...rrent output Current output 20mA analogue accessories Menu configuration 0 6 3 DANGER Open the housing of the TOPAX controller with disconnected power supply only Secure the power supply to prevent it...

Страница 15: ...54 55 61 51B 52B 71 72 73 74 75 76 77 78 39 40 41 43 44 45 46 42 35 95 96 A B A B pH value dosing pump PC interface Temperature pH value Redox 90 264 V AC Input module 5 fold CPRT ATE 5 fold Power sup...

Страница 16: ...12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 36 37 38 PE N L 33 34 92 93 94 87 88 89 90 91 86 81 82 83 84 85 56 57 58 59 60 51A 52A 53 54 55 61 51B 52B 71 72 73 74 75 76 77 78 39 40...

Страница 17: ...1 First set up and programming guidelines After correct installation switch on the power supply The TOPAX is programmed by the manufacturer based on customers specs before shipment and is supplied alo...

Страница 18: ...highlighted in grey To go back to the previous menu press the key back or 4 11 10 Menu 0 5 Completing the configuration The TOPAX moves immediately on to the configuration at the latest after menu 0 4...

Страница 19: ...hemical and physical correlations which are necessary to understand the behaviour of the measurement inputs The measurement inputs are processed for Disinfection pH value Redox potential Temperature A...

Страница 20: ...d using a two point calibration see section 12 1 5 Menu 1 2 Calibration on page 30 The physical measurement A measured on the measuring cell is shown on the display during calibration Reference value...

Страница 21: ...4 20 mA Whenever your TOPAX fea tures a lower value than 3 5 mA an alert message is displayed total chlorine sensor failure the alarm relay is switched 5 2 pH value measurement input 5 2 1 Technical d...

Страница 22: ...ble and the plug connectors If the zero point and transconductance values of the single rod measuring cell exceed the tolerances the system notifies the operator with a warning mes sage The measuring...

Страница 23: ...ixed buffer solution value The deviation should not exceed 10 approximately The plausibility of the transconductance measurement is monitored throughout the process After a reaction time approx 1 minu...

Страница 24: ...tal inputs are activated the level switches must be connected to the terminals respectively If no level switch is at tached the inputs are to be deactivated 7 Explanation of measuring values outputs 7...

Страница 25: ...ed current 0 4 20 mA or control outputs with pulse frequency The latter can be controlled as electronic optocoupler or relay outputs 8 Controller explanation 8 1 Definitions Term Definition actual val...

Страница 26: ...nd Ph close to the setpoints within tight limits e g for bathing the control ler must be adjusted to the controlled system This is achieved via the control parameters Xp for the proportional area Tn f...

Страница 27: ...f 20 of the control range 8 8 Manual mode Controllers can be manually adjusted If a controller is manually operated this will be indicated by a large M behind the Y in the status displays and the Y di...

Страница 28: ...nually adjusted to the actuator independently of the settings in the service menu recorder It is possible to configure the 0 20 mA controller output and a 4 20 mA recorder output It must be noticed th...

Страница 29: ...ecorder night mode service 2 mg l pH C 26 3 7 04 0 39 Fig 20 Menu 2 Extension of the main menu 12 1 3 Overview of the sub menus You can access all other sub menus through the main menu menu 1 and the...

Страница 30: ...electrode Menu 1 2 1 3 Encapsulated electrode Menu 1 2 1 pH value pH single rod measuring cell without specifications about slope rate Menu 1 2 2 pH single rod measuring cell with specifications abou...

Страница 31: ...1 4 2 Switching status of all outputs 1 4 3 Physical values and calibrated resistivity values of all inputs 1 4 4 used software version serial number hardware identification data input identification...

Страница 32: ...Xp value 2 1 2 3 2 Parameter Tn ON or OFF time 2 1 2 3 3 Parameter Tv ON or OFF time 2 1 2 4 Basic load ON or OFF time 2 1 2 5 Manual mode value ON OFF Select one of the two control units The device g...

Страница 33: ...digit of the limit value is highlighted in grey and can be ad justed Use the arrow keys to set up the value switch over between the individual positions alter numerical values Confirm the entry by pr...

Страница 34: ...s To select the menus press the arrow keys and Confirm the selection with OK 12 1 14 Menu 0 4 0 Configuration The configuration of TOPAX can be changed saved and loaded Existing configuration files ar...

Страница 35: ...rd Main menu menu 2 service password and service password In these menus you can block the various levels by using a password a 4 digit numerical code All information menus remain available The passwo...

Страница 36: ...a system name on the display Switching on off additional information on the standard display Setting the brightness of the display 12 1 29 Menu 5 18 Language Main menu menu 2 service language User lan...

Страница 37: ...OFF 14 Memory card The memory card transfers and stores language data and logfiles The data which is saved on the card can be transferred to a PC by means of a reading device To remove the memory card...

Страница 38: ...xceeded Activated configurable Min alarm Measurement below set level Activated configurable Y alarm Safety cutout see item 9 2 Activated configurable 15 3 Remedy Display Reason Remedy System start Sys...

Страница 39: ...t Disinfection Types of output Terminal connection clips Servomotor with potentiometer Motor Potentiometer with position feedback Servo motor without potentiometer Pulse frequency Optocoupler Pulse fr...

Страница 40: ...27 28 OK closed OK opened Not active 17 Device revision This operating manual applies to the following devices Device and model Revision status Software Version TOPAX 1 2013 1 25 onwards Language file...

Страница 41: ...t person Manufacturer order no Date of delivery Device type Serial number Nominal capacity nominal pressure Description of fault Service conditions of the device Point of use system designation Access...

Страница 42: ...ualquier modificaci n en la presente mercanc a sin nuestra previa autorizaci n esta declaraci n pierde su validez NL EU overeenstemmingsverklaring Ondergetekende Lutz Jesco GmbH Am Bostelberge 19 3090...

Страница 43: ...ata 5 H Hardware status 35 Housing 7 I Inputs module 8 Installation 7 K Key layout 17 L Language 36 Level dosing pump 24 Level inputs 24 Logbook 28 M Manual mode 27 Measurement input Disinfection 20 m...

Страница 44: ...com Internet www lutz jescoasia com sterreich Lutz Jesco GmbH Aredstra e 7 2 2544 Leobersdorf Austria Tel 43 2256 62180 Fax 43 2256 6218062 E Mail info lutz jesco at Internet www lutz jesco at Middle...

Отзывы: