![JDS Uniphase ORL-55 Series Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/jds-uniphase/orl-55-series/orl-55-series_operating-manual_3632969065.webp)
JDSU
ORL-55
61
8 R
EMOTE
C
ONTROL
:POW
:FETC
:AM
:AUTO
:FREQ?
Returns
• “1”, if an Auto-
modulation
frequency was detected on the
incoming signal.
• “0” otherwise
:POW
:FETC
:AM
:INT
:FREQ?
Returns
• “0”, if the incoming signal is not
modulated or Auto-
modulation
frequency is detected.
• the modulation frequency [Hz]
otherwise
Response type:
<NR1>
:FETC
:MEAS
:VAL?
Fetches the values measured by the
device depending on its mode:
POW: input power fetched
Resonse type: <NR3,NR3,NR3>
Response unit:
• Absolute display
(:POW:REF:STAT 0)
- in dBm for :POW:DISP:UNIT LOG
- in W for :POW:DISP:UNIT LIN
• Relative display
(:POW:REF:STAT 1) in dB
RET_LOSS: Return loss fetched
Response unit: dB
+9.91E37 overflow
9.91E37 underflow
The number of values fetched (1...3)
depends on the number of
wavelengths measured on the
incoming signal.
Command
string
Parameter type / Response type /
Unit / Info