![JDS Uniphase ORL-55 Series Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/jds-uniphase/orl-55-series/orl-55-series_operating-manual_3632969052.webp)
48
ORL-55
JDSU
7 M
AINTENANCE
7
M
AINTENANCE
Cleaning the test port
It is a good idea to check that the optical connections are
clean and to clean them if necessary before starting
measurements. Even very small dust particles on the
end surfaces of the plugs or in the test adapters can
adversely affect the accuracy of the measurement.
1. Switch off the device.
2. Remove the test adapter from the optical connection.
The plug end surface is now accessible.
3. Wipe off the plug end surface using a cotton bud
soaked in isopropanol. This cleaning method is very
effective and leaves no residues.
4. Blow out the test adapter with clean compressed air
(available in spray cans, e.g. Anti Dust Spray).
Note:
Cover the optical connections with the dust cap
whenever they are not in use. This prevents them from
getting dirty.
WARNING
Dangerous voltage and invisible
laser radiation
Maintenance or cleaning of the device
when it is connected up or operating
may damage the device or injure you.
!
Make sure that the device is switched
off and disconnected from all power
sources and optical radiation sources
before maintenance or cleaning.