background image

8

User ManUal

Westover fBP ProBe MICrosCoPe

JDSU's

 Westover  FBP-series 

probes  are  portable  video  microscopes  used  to  inspect  fiber  optic  connectivity. 

While most fiber microscopes are limited to inspecting 

male

 connectors, JDSU's FBP probe is designed to inspect 

both simplex and multi-fiber (ribbon) types of both 

male

 and 

female

 connectors, as well as optical devices such as 

transceivers. The probe is specially designed to fit and operate comfortably and easily in-hand, allowing the user 
to inspect hard-to-reach connectors that are installed on the backside of patch panels or inside hardware devices. 
This eliminates the need to disassemble hardware devices prior to inspection.

Focus Control

magnification Control

Inspection Tip

Barrel Assembly

standard Patch Cord Tips

standard Bulkhead Tips

JDSU’s  comprehensive  selection*  of  over  250  precision,  stainless-steel  fiber  inspection  tips  and  adapters  will 
inspect every connector and application. Our unique optics architecture and design provide true versatility and 
adaptability. These connector-specific and universal inspection tips are interchangeable, which allow the probe 
to interface with different types of fiber connectors. 

FBPT Tip Installation guide

Long reach Tips

60° Angled Tips

Barrel Assembly

ribbonDrive™ Tip

FBP PROBE OvERvIEW

FBPT INSPECTION TIPS

* Visit our web site for a complete list of 
inspection tips and adapters.

Содержание CleanBlast FCL-P1005

Страница 1: ...ber inspection and cleaning procedures documented in this manual are recommendations made by JDSU Please reference your company s process documents for standard tools and methods for your specific app...

Страница 2: ...s of Corning Incorporated Sony is a registered trademark of Sony Corporation Specifications terms and conditions are subject to change without notice All trademarks and registered trademarks are the p...

Страница 3: ...rs 15 Bulkhead Inspection 12 Patch Cord Inspection 13 Zones and Acceptance Criteria 14 Grading Process 14 Bulkhead Cleaning 16 Patch Cord Cleaning 17 Good Fiber Connection 18 Fiber Connections 18 Sing...

Страница 4: ...way to determine if fiber connectors are truly clean before mating them JDSU s video fiber inspection probe and CleanBlast system are used to quickly and easily inspect and clean connector end faces T...

Страница 5: ...key mechanism that aligns the fiber and prevents the rotation of ferrules of two mated connectors Ferrule Thin cylinder where the fiber is mounted that acts as the fiber alignment mechanism the end of...

Страница 6: ...creating unacceptable insertion loss and back reflection Furthermore these contamination particles can permanently damage the glass interface by digging into the glass and leaving pits when mated cau...

Страница 7: ...g debris into the fiber Cleaned Fiber Before Embedding Debris Cleaned Fiber with Embedded Debris After Mating and Multiple Cleanings If dirty connectors are not inspected and cleaned prior to mating d...

Страница 8: ...re devices This eliminates the need to disassemble hardware devices prior to inspection Focus Control Magnification Control Inspection Tip Barrel Assembly Standard Patch Cord Tips Standard Bulkhead Ti...

Страница 9: ...mination from the end face along with the solvent are then removed through the retrieval circuit The precise highly efficient non contact air solvent air stream blasts and removes loose debris with ne...

Страница 10: ...on Press this button to switch and select Brightness Contrast Color Tint and Reset modes Up Down Selector Press the Mode Button to select mode then press this button to adjust level of image quality e...

Страница 11: ...e To clean the SOCKET side 1 Seat the tip over one of the sockets 2 Pull the safety trigger to depress the safety release button 3 Apply slight pressure and push the handset into the insert until it s...

Страница 12: ...on page 26 4 Insert the probe into the bulkhead to inspect 5 Turn the focus control on the probe to focus the fiber image on the display 6 Determine whether clean or dirty If clean do not touch it and...

Страница 13: ...uide on page 26 4 Attach the patch cord to the probe to inspect 5 Turn the focus control on the probe to focus the fiber image on the display 6 Determine whether clean or dirty If clean do not touch i...

Страница 14: ...ector end face The inner most zones are more sensitive to contamination than the outer zones Acceptance criteria are a series of failure thresholds that define contamination limits for each zone A Cor...

Страница 15: ...one Name Diameter Defects Scratches A Core Zone 0 65 m 4 5 m none 5 m no limit 5 m 0 5 m B Cladding Zone 65 120 m no limit 2 m 5 from 2 5 m none 5 m no limit 5 m 0 5 m C Adhesive Zone 120 130 m no lim...

Страница 16: ...he bulkhead 4 Determine whether clean or dirty If clean do not touch it and CONNECT If dirty repeat CLEAN Note Embedded mated debris cannot be cleaned and bonded burned debris must be physically clean...

Страница 17: ...n green and press the run button to initiate cleaning 3 INSPECT the patch cord 4 Determine whether clean or dirty If clean do not touch it and CONNECT If dirty repeat CLEAN Note Embedded mated debris...

Страница 18: ...y throughout optical networks They give us the ability to re configure the network and provision services If contamination is present in the light path system performance will be degraded Always inspe...

Страница 19: ...oir must be refilled before operation can resume see page 20 for solvent refill procedures The reservoir will hold 225 ml 8 fl oz of solvent enough for at least 8 000 cleaning cycles Note Do not refil...

Страница 20: ...o disconnect the bottle and rotate the refill cap to original position and turn the system ON 6 Prime the system see page 16 Cap Lock 2 3 Solvent Refill Cap Assembly Cleaning Solvent Bottle see page 2...

Страница 21: ...th a flat head screwdriver unscrew and remove the 2 air filters and replace with new filters FCLP FA F1 6 Replace the water trap bowls tighten the drain valves and replace the release knobs 7 With a p...

Страница 22: ...need to be PRIMED Prime the system seepage16 Cleaning tip damaged or plugged Place the tip in electrostatic wash if problem persists replace tip D Unit operates but doesn t clean properly No cleaning...

Страница 23: ...pressure if stored shipped under elevated temperature Open closure slowly to vent pressure Contact 3M for the most current safety information and MSDS Reference 3M part number HFE 72DA The following f...

Страница 24: ...t enclosure has been opened other than as explained in this manual The equipment serial number has been altered erased or removed The equipment has been misused neglected or accidentally damaged This...

Страница 25: ...ning tip 6 4 in LCD EU power FCL P2100 CleanBlast portable with 5 ft umbilical 90 bulkhead handset USB 1 1 output universal 2 5 mm cleaning tip FCL P2102 CleanBlast portable with 5 ft umbilical 90 bul...

Страница 26: ...FC UPC connectors through a bulkhead barrel assembly required ST UPC FBPT ST Inspect ST UPC connectors through a bulkhead barrel assembly required LC UPC FBPT LC L Inspect LC UPC connectors through a...

Страница 27: ...ors through a bulkhead ST UPC FCLT U25 Clean 2 5 mm UPC and APC connectors through a bulkhead LC UPC FCLT U12 Clean 1 25 mm UPC and APC connectors through a bulkhead FCLT U12 MA FCLT U12 Clean 1 25 mm...

Страница 28: ...ssumes all risks and liability whatsoever in connection with the use of a product or its applications JDSU reserves the right to change at any time without notice the design specifications function fi...

Отзывы: