17
CleanBlast – PortaBle
Clean:
PATCH CORD
CLEAN
CleanBlast
Advanced Fiber End Face Cleaning System
1.
Install the appropriate cleaning tip and
adapter
(fCLT-U25 and fCLT-U25-MA shown)
to the CleanBlast handset.
Note: Patch cord cleaning
requires a mating
adapter (see
Appendix B: Cleaning Tips &
Adapters Guide
on page 27).
Cleaning Tip
(FCLT-U25)
Adapter
(FCLT-U25-MA)
1
3
4
2.
Attach the patch cord to the handset, push
the connector into the handset to disengage
safety mechanism
(the POWER/READY light
will turn green),
and press the
run button
to
initiate cleaning.
3.
INSPECT
the patch cord.
4.
Determine whether
clean
or
dirty
.
If clean,
do not touch it and
CONNECT
.
If dirty,
repeat
CLEAN.
Note:
Embedded/mated debris cannot be
cleaned, and bonded/burned debris must be
physically cleaned before using CleanBlast.
run Button
2
PRImE
run Button
Prime Button
(press 2 times)
PRImE / PuRgE
the nozzle to clear the solvent line if
you are using the system for the first time, after refill,
or if the system has been powered oFF for more than
4 hours.
A.
Push the
prime button
quickly
2 times.
B
.
Aim the tip away from any surfaces or objects.
C.
Press the
run button
on the handset.
D.
Repeat steps A through C
at least 2 more times.