© 2016 JCM American Corporation
1 7
November, 2016
9. Follow the onscreen prompts and click the
Start
button (Figure 9
a
) to begin the
DBV-500 Validation Sensor Calibration.
10.When prompted to
Change Serial Number?
, click
Yes
if the Serial Number
recorded on the CPU Circuit Board needs to be changed; otherwise, click
No
to
complete Calibration.
11.When the Sensor Calibration Screen appears (Figure 10), click the
Start
button
a
) to save the Calibration Settings to EEPROM.
NOTE: KS-091 Reference Paper must be used for Calibration of the DBV-500.
a
Figure 9
DBV-500 Sensor Calibration Function
a
b
Figure 10
DBV-500 Sensor Calibration Results