Section 10. Diagnostics
Use can see what happen in flowmeter through diagnostic functions.
FREQUENCY DIV is frequency fuctions
RISC is distance from impulse signal to receive signal
FULSE COUNT is numbers of pulse. 5 is default.
GAIN LEVEL is amplitude level of signal. The number is automatically set by flowmeter.
HOLD FLOW is functions to match with remote indicator.
When HOLD FLOW is 0, Xonic 100 send 4mA output.
When HOLD FLOW is 1000 (max span flow), Xonic 100 send 20mA output.