background image

26 

 

Section 9. DATALOGGER 
 

 

 

 

 

 

Please set TIME SET after installation. 
Please set RS-232C parameters to communicate with your computer. 
 

Содержание XONIC-100P

Страница 1: ...1 www jain co kr 82 2 856 4114 USER MANUAL XONIC 100P PORTABLE FLOWMETER ...

Страница 2: ...sducer Installation Place Section 1 GENERAL 1 1SITE NAME 7 1 2 PASSWORD 7 1 3 METER TYPE 8 1 4 APPLICATION 9 1 5 DISPLAY 9 Section 2 CH SELECT 10 Section 3 PIPE data 11 Section 4 LIQUID type 13 Section 5 INSTALLATION 14 Section 6 OPERATION 18 Section 7 FLOW 21 Section 8 IN OUTPUT 24 Section 9 DATALOGGER 26 Section 10 Diagnostics 27 DRAWINGS 28 ...

Страница 3: ...3 Delivery Items Flow Computer Transducer Mounting Track Cable Easy Track Magnetic Track ...

Страница 4: ... 7D Down stream 3D dual path Transducer 40 120 Flow Total Velocity Delta T Oscilloscope Shape Temperature Flow Computer 20 60 4 Accuracy 1 0 single path 0 5 dual path Flow Velocity 0 02 20 0 m s Turn Down Ratio 1000 1 Repeatability 0 25 Required Straight Run Xonic 100 use inner batteries and AC 85 264V 50 60Hz power adapter So please consider about safety and any damage to flow computer during ins...

Страница 5: ...play Site Name user must input site name for communication Type single ch means single path channel Flow Unit user selected flow unit Flow measured flow Total total flow Total Unit total flow unit Keys Functions Press when start MENU Accept Key Data Clear 0 9 numeric Use when input numbers Use when input Function Key Direction keys Negative input ...

Страница 6: ...allation place which has enough straight pipe runs 10D for upstream and 5D for down stream For example if pipe dameter is 1000mm then please find 10m straight run for up stream and 5 straight run for down stream Find 10D for Upstream and 5D for Downstream ...

Страница 7: ...or D and input data User can input alphabet character by pressing keys and to complete Press UP DN key until cursor show J Press UP DN key until cursor show A Press UP DN key until cursor show I Press UP DN key until cursor show N The cursor will move to next menu If user want to finish menu press User can use alphabet and space characters 1 2 PASSWORD User must input password to secure Xonic 100 ...

Страница 8: ...ows FLOWMETER Clamp On flowmeters for single path dual path channel VEL NORMAL MODE Velocity Meter mode for m s base VEL TIME MODE Velocity Meter mode for time base Single Path V Method Installation Single Path Z Method Installation Dual Path V Method Installation ...

Страница 9: ...nd dual path channel mode User can buy only single path type and can add one more channel if need 1 5 DISPLAY Xonic 100 display only channel 1 or channel 2 or two channel simutanously User can select by cursor Single Path Display Dual Path Display ...

Страница 10: ... one pipe Dual channel means user will use 2 pair transducers for two pipes In case of dual channel flowmeter measure two pipes simultaneously Also user can select single path or dual path Single Path means user will use 1 pair transducer for one pipe Dual Path means user will use 2 pair transducers for one pipe to increase accuracy ...

Страница 11: ...11 Section 3 input PIPE DATA 3 1 Flange Tye Just select NONE 3 2 Pipe Unit select METRIC or US INCH 3 3 Pipe Material select pipe material from list 3 4 Pipe Sonic Vs N A ...

Страница 12: ...Diameter Input pipe diameter using numeric keys 3 6 Pipe Thickness Input pipe thickness using numeric keys 3 7 Lining Material select from list 3 8 Lining Sonic Vs N A 3 9 Lining Thickness input using numeric keys ...

Страница 13: ...wn user must input 3 VISCOSITY and 4 DENSITY manually 4 1 LIQUID TYPE Please select from liquid list 4 1 SONIC VELOCITY User do not need anything flowmeter automatically display sound speed of selected liquid 4 2 VISCOSITY User do not need anything flowmeter automatically display viscosity of selected liquid 4 3 DENSITY User do not need anything flowmeter automatically display density of selected ...

Страница 14: ...ransducers and Xonic 100 will automatically recommend proper transducers for the site If Xonic 100 recommend size XLD type then usr must have XLD type transducers Choose from the list transducers you have 5 2 Select SAMPLING CLK Xonic 100 automatically select SAMPLING CLOCK so user do not need change 5 3 MOUNTING TYPE Normally CLAMP ON V is better for most applications V method is more accurate an...

Страница 15: ...lowmeter can not work Also in case of water is not clean for example waste water please use Z MODE V mode installation This Z MODE make ultrasonic signal more strong than V MODE 5 4 INSTALL INFO 5 4 1 Finding Installation Position Please find enough straight run pipe position Normally clamp on ultrasonic flowmeter need 10 Upstream and 5 Downstream straight pipe run Please find such position ...

Страница 16: ...e install position flat and clean Please peel off pipe coating and Use abrasive to clean the area 5 4 4 Install Mounting Track Install mounting track onto the pipe with stainless steel strap Fix it tightly 센서고정용트랙 트랙고정용 스트랩 Space 센서의 설치거리 5 4 5 Install Transducers onto Pipe Apply couplant onto bottom of transducers and locate transducer into mounting track Make transducers clamp on pipe by tighten...

Страница 17: ...be below 1500 High gain means low ultrasonic signal So if gain is over 1500 please check pipe size again and re install transducers after clean pipe 3 Signal Shape must be high in the middle area If signal shape is not high in the middle area please check pipe size pipe material etc 5 6 QUICK INSTALL Similar to auto installation Do not use this menu at first installation This menu is useful when a...

Страница 18: ...OW LIMIT Normally this value is 0 6 3 DEAD ZONE DEAD ZONE means the flow than can be disregarded If pipe is big so very small flow in meaningless then use this menu Normally flowmeter make the value automatically 6 4 FLOW AVERAGE TIME Default value is 5 seconds This means display flow is average flow during 5 seconds 6 5 TOTAL FLOW SET If user need to change flow total can change by keypad 6 6 ALA...

Страница 19: ...d want more accurate measurement use can select calibration menu Move cursor to CALIBRATION METHOD 1 NO CALIBRATION is no calibration If do not affect any calibration to the flow 2 MULTI POINTS is multi point calibration menu User can test flow from minimum to max flow And can input each test points to flowmeter VIEW user can see each input points ADD user can add test points ...

Страница 20: ...libration with calibration factor If flow is 100 and Kc is 1 0 then flow became 100 If flow is 100 and Kc is 1 01 then flow became 101 If flow is 100 and Kc is 0 09 then flow became 99 9 6 8 ENABLE AGC AGC is Automatic Gain Control function Enable is default ...

Страница 21: ...2 FLOW TIME UNIT User can select time unit from list 7 3 FLOW RESOLUTION User can select decimal points from list XXXX means 10 for flow XXX X means 10 1 for flow XX XX means 10 12 for flow X XXX means 10 123 for flow 7 4 FLOW SCALE In case of big flow user can select KILO menu ...

Страница 22: ...it 7 6 TOTAL RESOLUTION Choose from the list 7 7 TOTAL SCALE User can select KILO for big flow total 7 8 BATCH TOTAL Batch total means relay will be on per each batch total If flow unit is CUBIC METER then 1 0 means 1 pulse per 1 CUBIC METER If 0 1 batch total it means 1 pulse per 0 1 CUBIC METER ...

Страница 23: ...f positive flow is 100 and negative flow is 10 then total is 90 POSITIVE TOTAL means flowmeter will only totalize positive flow If positive flow is 100 and negative flow is 10 then total is 100 NEGATIVE TOTAL means flowmeter will only totalizer negative flow If positive flow is 100 and negative flow is 10 then total is 10 ...

Страница 24: ...output User can assign output data and set range User can assign FLOW or VELOCITY to ANALOG OUT 1 In most case flow is assign to ANALOG OUT 1 User can set ANALOG OUT 1 min and max span value If flow max is 1000 SPAN MAX is 1000 If flow min is 0 SPAN MIN is 0 8 2 ANALOG OUT 2 Same as ANALOG OUT 1 ...

Страница 25: ...ATCH TOTAL 8 4 RELAY OUT 2 Same as RELAY OUT 1 8 5 ANALOG IN 1 In case of user want to see pressure temperature use can use this function Just set MIN and MAX input SPAN then flowmeter send the ANALOG INPUT data through RS 232C If pressure transmitter range is from 0 to 10Kg cm2 then MIN INPUT SPAN is 0 MAX INPUT SPAN is 10 8 6 ANALOG IN 2 Same as ANALOG IN 1 ...

Страница 26: ...26 Section 9 DATALOGGER Please set TIME SET after installation Please set RS 232C parameters to communicate with your computer ...

Страница 27: ...istance from impulse signal to receive signal FULSE COUNT is numbers of pulse 5 is default GAIN LEVEL is amplitude level of signal The number is automatically set by flowmeter HOLD FLOW is functions to match with remote indicator When HOLD FLOW is 0 Xonic 100 send 4mA output When HOLD FLOW is 1000 max span flow Xonic 100 send 20mA output ...

Страница 28: ...28 DRAWINGS ...

Страница 29: ...29 POWER ...

Страница 30: ...30 ...

Страница 31: ...31 CIRCULAR CONNECTOR ...

Страница 32: ...32 FLOW TRANSDUCER CABLE CONNECTOR ...

Страница 33: ...33 TRANSDUCERS INSTALLATION 102 TP AND MOUNTING TRACKS DIRECT MODE DIRECT MODE INSTALLATION STRAP GUIDE MOUNTING TRACK SERIES CUSTOMER S PIPE FLOW ...

Страница 34: ...34 REFLECT MODE INSTALLATION 102 TP TO FLOW COMPUTER UP STRAP GUIDE TRANSDUCER MOUNTING TRACK SERIES CUSTOMER S PIPE CABLE FLOW TRANSDUCERS INSTALLATION AND MOUNTING TRACKS REFLECT MODE ...

Страница 35: ...35 INSTALLATION OUTLINE 102 S SF ADJUSTABLE MOUNTING STRAP ...

Страница 36: ...36 USE OF COUPLANT COMPOUND TRANSDUCERS INSTALLATION 102 C COUPANT APPLY A THIN BAND OF COUPLANT TP BASE OF EACH TRANSDUCER BEFORE INSTALLING ON PIPE ...

Страница 37: ...37 CABLE CONNECTING KIT CABLE CONNECTING 1 2 3 4 5 6 7 ...

Отзывы: