6.7 CALIBRATION
In case of user has calibration laboratory or other laboratory test Xonic 100, then they can use this
menu for best accuracy. If flowmeter has some accuracy and want more accurate measurement,
use can select calibration menu.
Move cursor to CALIBRATION METHOD.
1) NO CALIBRATION is no calibration. If do not affect any calibration to the flow.
2) MULTI-POINTS is multi-point calibration menu. User can test flow from minimum to max flow.
And can input each test points to flowmeter.
VIEW user can see each input points
ADD user can add test points