BM-500
CL / BB-500
CL
- 30 -
Command Name
Format
Parameter
Remarks
D -Image Format
1
Bit Allocation
BA
=[Param.]<CR><LF>
BA?<CR><LF>
0=8bit, 1=10bit, 2=12bit
2
Scan Format
SC
=[Param.]<CR><LF>
SC? <CR><LF>
0=Full Frame
1=2/3 Partial
2=1/2 Partial
3=1/4 Partial
4=1/8 Partial
5=Decimation (Draft)
3
V-Binning
VB
=[Param.]<CR><LF>
VB?<CR><LF>
0=OFF
1=On
Only for BM-500CL
4
Variable partial
scan
PRGP
=[Param.]<CR><LF>
PRGP?<CR><LF>
0=OFF
1=ON
For Programmable
partial
5
Start line
STL
=[Param.]<CR><LF>
STL?<CR><LF>
2 to 2058
For Programmable
partial
6
End line
ETL
=[Param.]<CR><LF>
ETL?<CR><LF>
2 to 2058
For Programmable
partial
E - Gain, Black and signal settings
1
Gain Level
GA
=[Param.]<CR><LF>
GA?<CR><LF>
-84 to 336
2
Gain Adjust 2
GJUT2
=[Param.]<CR><LF>
GJUT2?<CR><LF>
-84 to 84
3
Gain Mode
AGC
=[Param.]<CR><LF>
AGC?<CR><LF>
0= Manual Gain Control
1=Auto Gain Control
4
AGC Reference
AGCF
=[Param.]<CR><LF>
AGCF?<CR><LF>
0 to 8192
5
Black Level
BL
=[Param.]<CR><LF>
BL?<CR><LF>
255 to 767
6
User Black
set up1
BLS
=[Param.]<CR><LF>
BLS?<CR><LF>
-511 to 511
For L ch
7
User Black
set up 2
BLS2
=[Param.]<CR><LF>
BLS2?<CR><LF>
-511 to 511
For R ch
8
Black fine
BLF2
=[Param.]<CR><LF>
BLF2?<CR><LF>
-255 to 256
For R ch
F – AWB, LUT and others
1
AWB mode
AWBM
=[Param.]<CR><LF>
AWBM?<CR><LF>
0=One push 1=3200K
2=4600K 3=5600K
Only for BB-500CL
2
One Push AWB
AWB
=[Param.]<CR><LF>
1=Run
Only for BB-500CL
3
One Push AWA
AWA
=[Param.]<CR><LF>
1=Run
Ch-to-ch gain
4
One Push ABA
ABA
=[Param.]<CR><LF>
ABA?<CR><LF>
1=Run
Ch-to-ch black
level