BM-500
CL / BB-500
CL
- 28 -
7.3.7.
Trigger polarity. TP=0, TP=1.
The active trigger polarity is normal low (TP=0). It can be invert it to active high (TP=1).
7.3.8.
Gain level. GA=-84 t336.
GA=0 is 0dB gain, which is normal working point. The range is from -3 dB to +12 dB.
7.3.9.
Black level. BL=0 through BL=1023.
Black level (or set-up level) will set the video level for black. Factory setting is 128 LSB for 12 bit,
32 LSB for 10bit or 8 LSB for 8bit.
7.4.
Save and Load Functions.
The following commands are for store and load camera settings in the camera EEPROM.
Load settings.
LD.
This command will load previous stored settings to the camera. 3 user settings can be stored in
the camera EEPROM. 1 factory setting is also stored in the camera. The settings stored in the last
used user area is used as default settings at power up.
Save Settings.
SA.
This command will store the actual camera settings to 1 of the 3 user area in the camera EEPROM.
EEPROM Area.
EA.
If received, the camera will return
the last used user area number.