AM-1600GE / AB-1600GE
61
11.2
Specification table
Specifications
AM-1600GE
AB-1600GE
Scanning system
Progressive scan
Frame rate full frame
3.04 frames/sec. Progressive (3327 lines/frame)
Pixel clock
30 MHz
Line frequency
10.135 kHz (1H = 98.66 µs)
(2960 pixel clock/line L ch and 2960 pixel clock / line R ch)
CCD sensor
43.3mm diagonal Monochrome
KAI-16000-AXA
43.3mm diagonal Bayer Color
KAI-16000-CXA
Sensing area
36.05 (h) x 24.035 (v) mm 43.3mm diagonal
Cell size
7.4 (h) x 7.4 (v)
μ
m
Active pixels
4872 (h) x3248 (v)
Pixels in video output.
Full scan
1/2 partial scan
1/4 partial scan
Variable partial scan
4872 (h) x 3248 (v) 3.04 fps. H = 10.135 kHz
4872(h) x 1372 (v) 5.51 fps H= 10.135 kHz
4872 (h) x 1028 (v) 9.25 fps. H = 10.135 kHz
Start line from 1 to 2448 by 1 line unit for AM-1600GE, 2 lines units for
AB-1600GE , height(lines) from 800 to 3248
Vertical binning mode
2x (5.63 fps )
n/a
Sensor sensitivity
Output : 30
μ
V/e
Saturation: 30,000 electrons
Sensitivity on sensor (minimum)
0.02 Lux (Max. gain, Shutter OFF,
50% video )
0.35 Lux (Max. gain, Shutter OFF,50%
Green, w/IR cut filter)
S/N ratio
More than 56 dB (0dB gain)
More than 56 dB (Green, 0dB gain)
Digital video output
GigE Vision Compliant
Mono8,Mono10,Mono10_Packed
Mono 12, Mono12_Packed
GigE Vision Compliant
BAyGR8, BAYGR10
BAYGR12
White balance
n/a
Manual/One push
Gain
Manual Master gain: -3dB to +12dB
Individual gain : -6dB to +6dB (AB-1600GE only)
AGC : -3 to +12 dB
Black level
Default setting: 32LSB(10 bits output), Range : 0 LSB to 128 LSB
Gamma/Knee LUT
(
γ
=1.0(OFF),
γ
=0.45(ON) )
Blemish correction
ON / OFF, Calibration
n/a
Shading compensation
Provided
n/a
L/R channel balance
provided
Test pattern generator
Provided
Synchronization
Internal X-tal
GPIO Module
Input/output switch
Clock generator ( One)
Pulse generators ( Two)
Configurable 13-in / 10-out switch
12-bit counter based on Pixel clock
19-bit counter programmable for length, start point, stop point, repeat
Hardware trigger modes
Edge Pre-Select , Pulse Width Control, Frame Delay and Sequence
OB area transfer mode
ON / OFF
Event message
Exposure start, Exposure end, Trigger IN, Video start, Video end