Sterilization temperature
121°C
134°C
Sterilization time
15 to 50 minutes
4 to 40 minutes
Drying time
1 to 30 minutes
1 to 30 minutes
Vacuum number
0, 1 or 3
0, 1 or 3
6.4.3 Set language
Press UP or DOWN button to choose language, and then press START/STOP to save and exit.
6.4.4 Program 11 USER settings
Program No. 11 “User-defined” is used to achieve a perfect sterilization by adjusting to a special setting. The process of the “User-
defined” program is as follows:
- Press the PROG key (4) to enter program 11, then press PROG key (4) again to select the value you wish to modify: sterilization
temperature, sterilization time, drying time and number of vacuums.
- Press the UP (2) or DOWN (3) keys to change the value of each parameter.
- Press the START key (5) to confirm and START (5) again to start the cycle.
Steam Sterilizer Operation Manual
No. ASS0010
REV-D
Page 25 of 48
5.2.2.3: Language Setting
Press
or
button to choose language, and then press
to save and exit.
5.2.3 Special Sterilization Setting
Program No. 11
“
User-defined
” is used for setting up the special data in order to get
perfect sterilization effect. “
User-defined
” program operation process.
Press
to enter
the “
User-defined
” program,
then press
to switch items:
sterilization temperature, sterilization time, drying time, times of vacuum
Press
or
button to set value of each item.
Press
button to confirm, press again to start the cycle.
11
User-defined
134
℃
4min
03min 3time
SELECT
11
User-defined
134
℃
4min
03min 3time
SELECT
LANGUAGE
1. English
5. P
усский
2. Español
6.
Italiano
3. Portugês
7.
Francese
4.
中文
CAUTION when removing the material from the autoclave.
Risk of burns due to contact with hot surfaces.
The material inside the autoclave could be at a temperature higher than 100ºC.
Use the handle provided for trays handling.