INSTRUCTION MANUAL
CODE 80416 REV D
12/2023
Page: 36
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: http://www.grupo-selecta.com
4.1 Autoclaves type «B»
Autester ST DRY PV
autoclaves allow sterilization of:
• Solid loads resistant to temperature: instrumental, metallic.
• Solid loads sensitive to temperature: plastics
• Solid packed load.
• Porous load: textiles
• Wrapped porous load.
• Catheters, tubes and cavities devices.
• Liquids.
4.2 Technical specifications
Model
4001738 4001739
Supply voltage:
220V, 50Hz
220V, 50Hz
Power:
1500VA
1700VA
Fuse:
F12AL
F12AL
Working temperature:
5-40ºC
5-40ºC
Noise:
< 50dBA
< 50dBA
Maximum weight per tray:
1 Kg
1 Kg
Maximum heat radiation (from 20 to 26ºC): < 2000J
< 2000J
Chamber material:
Stainless steel
Stainless steel
Maximum working pressure:
2.5bar
2.5bar
Minimum working pressure:
-0.9bar
-0.9bar
Maximum temperature:
145ºC
145ºC
Chamber diameter measurements:
245x360mm
245x470mm
Water volume per cycle:
0.16L
0.18L
Water tank volume:
3.5L
4L
4.3 Sterilization cycle
1-2 pre-heating
2-3 pre-vacuum
3-4 sterilizing
4-5 air-discharging
5-6 drying
6-7 stabilizing
1-7 entire duration
Sterilization cycles types
Type B:
The sterilization of all wrapped or non-wrapped, solid, hollow load products type A and porous pro-
ducts as represented by the test loads in this standard.
Type S:
The sterilization of products as specified by the manufacturer of the sterilizer including non-wrapped
solid products and at least one of the following: porous products, small porous items, hollow load
products type A, hollow load products B, single wrapped products, multiple-layer wrapped products.
NOTE 1
The description identifies ranges of products and test loads.
NOTE 2
Non wrapped sterilized instruments are intended either for immediate use or for non sterile
storage, transport and application (e.g. to prevent cross infection).