If the water shortage alarmed during the sterilizing, you should not worry about it, the water will not influence this process.
The water that had been left in the tank is adequate for the operating cycle, but you need to add water in time, so as to
ensure the next operation.
CAUTION: USE DISTILLED WATER ONLY TO EXTEND THE LIFE TIME OF THE MACHINE. DO NOT TILT
THE STERILIZER WHEN THE RESERVOIR IS FULL OF WATER.
Alarming if Used Water Reservoir is full
The LCD screen shows WATER DRAINING ARROW during the sterilizing, which means that the used water reservoir needs
to be drained out.
Connect the water tube to the machine left lower corner inlet connector as right picture arrow marked, please see picture.
Used water outlet
Generally, the maximum temperature of drained water should be under 70ºC. If it is higher, you need to check whether
the fan operates normally, or contact the local distributor immediately, we will offer our best service to you in no time.
Steam Sterilizer Operation Manual
No. ASS0010
REV-D
Page 29 of 48
so you need to add distilled water in until the lowest required level can be reached.
You can fill in water automatically or by manual on the top of the machine,
Automatically: Connect the water tube with the machine left lower corner inlet connector as
right picture arrow marked, please see the picture, and press
button for 3 seconds;
by manual: open up the top cover, to fill in water according to the pic 3, stop filling in water
again when you hear beep sound.
If the water shortage alarmed during the sterilizing, you should not worry about it, the water will
not influence this process. The water that had been left in the tank is adequate for the
operating cycle, but you need to add water in time, so as to ensure the next operation.
CAUTION USE DISTILLED WATER ONLY TO EXTEND THE LIFE TIME OF THE
MACHINE.
DO NOT TILT THE STERILIZER WHEN THE RESERVOIR IS FULL OF WATER.
6.3 Alarming if Used Water Reservoir is full
The LCD screen shows WATER DRAINING ARROW during the sterilizing, which means that
the used water reservoir needs to be drained out.
Connect the water tube to the machine left lower corner inlet connector as right picture arrow
marked, please see picture.
OPEN THE DOOR BEFORE WORKING
CLOSE THE DOOR
Water Filling Port
Water Filling Port
Used water tank full alarm
Steam Sterilizer Operation Manual
No. ASS0010
REV-D
Page 29 of 48
so you need to add distilled water in until the lowest required level can be reached.
You can fill in water automatically or by manual on the top of the machine,
Automatically: Connect the water tube with the machine left lower corner inlet connector as
right picture arrow marked, please see the picture, and press
button for 3 seconds;
by manual: open up the top cover, to fill in water according to the pic 3, stop filling in water
again when you hear beep sound.
If the water shortage alarmed during the sterilizing, you should not worry about it, the water will
not influence this process. The water that had been left in the tank is adequate for the
operating cycle, but you need to add water in time, so as to ensure the next operation.
CAUTION USE DISTILLED WATER ONLY TO EXTEND THE LIFE TIME OF THE
MACHINE.
DO NOT TILT THE STERILIZER WHEN THE RESERVOIR IS FULL OF WATER.
6.3 Alarming if Used Water Reservoir is full
The LCD screen shows WATER DRAINING ARROW during the sterilizing, which means that
the used water reservoir needs to be drained out.
Connect the water tube to the machine left lower corner inlet connector as right picture arrow
marked, please see picture.
OPEN THE DOOR BEFORE WORKING
CLOSE THE DOOR
Water Filling Port
Water Filling Port
Used water tank full alarm
6.2 Program selection
• Turn on main switch.
• Press PROG key (4) to access the options PROGRAM and CONFIGURATION.
• Press the UP key (2) to access the preset programs.
• Select the program with keys (2) and (3).
• Confirm by pressing PROG key (4).
Steam Sterilizer Operation Manual
No. ASS0010
REV-D
Page 21 of 48
5.2 Menu
5.2.1 Main menu
When you open the power switch you will see the operation information and last program on
the window.
press
button to enter the choosing menu,
WINDOW
DESCRIPTION
EMERGENCY: 134
℃
/5min, it is a fast working
cycle, one time vacuum; working temperature
134
℃
; Working pressure 2.1bar; sterilization cycle
time 4 minutes; and drying time 8 minutes.
SOLID: 134
℃
/5min, it is used for nude and solid
instruments, one time vacuum, working
temperature 134
℃
, working pressure 2.1bar,
sterilization cycle time 5 minutes, drying time 8
minutes.
2 SOLID
134
℃
05min
08min 1time
SELECT
1 EMERGENCY
134
℃
04min
08min 1time
SELECT
PROGRAM
SET
EMERGENCY
134
℃
-04min-08min-1
(00015)
PRINTER OFF
12-05-2012 10:28:36
Open the door before working
CLOSE THE DOOR