ModBox VNA
–
User’s Manual
Ref. Ed. B
–
12/05/2020
30
4.3
AN MBC manual calibration:
Step Action
1.
Check that a laser is turned on.
2.
Go into
Configuration (Settings gear) > MBC > Feedback
tab and set the
MBC in manual mode. Find the maximum light on the output modulator by
changing DC bias value.
3.
After that go to PD Gain tab, adjust PDL gain to obtain 2V for VPDL and adjust
PDM gain to obtain 4V for VPDM.
4.
Go back to Feedback and set MBC in AUTO mode. Modulator locking point
will be QUAD.