B-H ANALYZER SY-8218
/ SY-8219
51
Table 7-1 Error/ warning messages
Displayed messages
Causes
Remedies
W19
: H range over at 1st excitation!
Overflow occurs in the exciting current
measurement range during approximate value
excitation.
If Hm is specified in Fixed Parameter,
confirm
that
numeric values and units of Le and N
1
are correct.
W20
: B range over at 1st excitation!
Overflow occurs in the induced voltage
measurement range during approximate value
excitation.
If Bm is specified in Fixed Parameter, confirm that
numeric numbers and units of Ae and N
2
are
correct.
W23
: SST serial communication
is abnormal!
Serial communication with SY-956 (SST) is
abnormal.
Restart this product and SY-956.
When this phenomenon happens frequently,
contact Iwatsu service center to tell the error
message No.
W24
: DCT serial communication
is abnormal!
Serial communication with SY-960 (DCT) is
abnormal.
Restart this product, SY-960 and SY-961.
When this phenomenon happens frequently,
contact Iwatsu service center to tell the error
message No.
W25
:
DCS serial communication
is abnormal!
Serial communication with SY-961 (DCS) is
abnormal.
Restart this product, SY-960 and SY-961.
When this phenomenon happens frequently,
contact Iwatsu service center to tell the error
message No.
W30
: Setting condition error! V2
200V
In measurement condition, induced voltage
exceeds 200[V].
Confirm that values and units of N2, Ae, freq. are proper.
Confirm that values and units of Bm specified in
Fixed Parameter is proper.
Reduce N2 of the sample.
W32
: Setting condition error! I1
6A
In measurement condition, exciting current
exceeds 6[A].
Confirm that values and units of N
1
and Le are proper.
Confirm that value of Hm specified in Fixed
Parameter is proper.
Increase N
1
of the sample.
W34
: Setting condition error!
Yoke V3 or V4 > 200V
It is forecast that the output voltage of the third
or fourth turns of the measurement yoke
exceeds 200V when 2nd is measured by the
Single Seat Measurement.
Confirm whether the value specified with Fixed
Parameter is too large.
Confirm whether the value of Ae is too large.
W35
: Setting condition error!
⊿
VL >
400V
It is a measurement condition for the increment
voltage to exceed 400V
p-p
by the DC bias
measurement.
Confirm that values and units of N2, Ae, freq. are proper.
Confirm that values and units of Bm specified in
Fixed Parameter is proper.
Reduce N2 of the sample.
W36
: Setting condition error!
⊿
IL >
12A
It is a measurement condition for the increment
current to exceed 12A
p-p
by the DC bias
measurement.
Confirm that values and units of N
1
and Le are proper.
Confirm that value of Hm specified in Fixed
Parameter is proper.
Increase N
1
of the sample.
W41
: OSC output under!
0V
OSC output is tried to be set under 0V during
approximate value excitation.
Confirm whether the value specified with Fixed
Parameter is too small.
Reduce the gain of the power amplifier.
W42
: Shortage of the exciting current!
The exciting current is insufficient by asymptote
excitation.
Confirm whether the sample is connected or
disconnected.
W43
: OSC output under!
≦
0V
OSC output is tried to be set under 0V during
asymptote excitation.
Confirm whether the output of the power amplifier
is proper.
Reduce the gain of the power amplifier.
W50
: Measured V2m over!
> 200V
Induced voltage measured exceeds 200[V].
Confirm that values and units of N
2
, Ae, and freq.
are proper.
Confirm that value and unit of Bm specified in
Fixed Parameter are proper.
Reduce N
2
of the sample.