B-H ANALYZER SY-8218
/ SY-8219
33
■
6.4 Setting of measurement condition
●
The measurement condition is set.
①
While referring to Table 6-3, move the edition cursor to Edit Box of the measurement condition to be entered on
Excited Conditions
and
press to determine the selection.
②
When Popup List opens, move the edition cursor to the item to be set and press to determine the setting.
➂
When Popup Edit Box opens, use the ten key to enter the number and press . Or move the edition cursor to the number to be
changed and press
. Or, rotate
to change the number. Press to determine the setting.
Table 6-3 Measurement condition
[Explanatory note]
Bold
: Factory settings
6-2
:
μ
mode cannot select a square wave. This product generates the symmetry square wave with
Duty
50
.
Items
Selectable items/ input rules
Description
Function
Sine
Sine wave
Selects a type of excitation signal.
Pulse
6-2
Square wave
Calc.
Normal
Selects calculation mode.
μ
Frequency
Sine wave
SY-8218
10Hz to 10MHz
Selects a frequency of excitation signal.
SY-8219
10Hz to 1MHz
100
Square
wave
10Hz to 1MHz
Fixed Parameter
(Target parameter)
Bm
(Max. magnetic flux density)
0.001 to
99999
6-3
Selects and sets a target parameter value for
measurement.
Hm
(Max. magnetic field)
I1m (
Max.
excitation current)
V2m (
Max. induced voltage)
1
Tolerance
0.1~100%
6-3
Sets the tolerance
X
. The tolerance
X
is used for the
criteria to finish measurement; i.e.
X
[%] is the ratio of
an actual value to a target parameter setting value
and if X is reached, measurement finishes.
1
Содержание SY-8218
Страница 1: ...Instruction Manual Analyzer SY 8218 SY 8219...
Страница 2: ...2011_2017 IWATSU ELECTRIC CO LTD All rights reserved...
Страница 57: ...B H ANALYZER SY 8218 SY 8219 45 Fig 6 15 Reference function in time axis graph...
Страница 77: ...SY 8218 SY 8219...