B-H ANALYZER SY-8218
/ SY-8219
5
■
1.3 Cautions for transportation
●
When transporting this product, remove the measurement POD from the main unit and use the packing material attached at the time of
purchase or equivalent.
The case where a failure or damage occurs when transporting this product with the measurement POD attached to the main unit
shall not be
guaranteed
.
●
Remove the measurement POD as follows:
①
Loose two clamp knobs on the front of the measurement POD by rotating them counterclockwise.
②
Apply both hands on both sides of the measurement POD and pull the measurement POD to your side horizontally along the guide rails.
Never use excessive power to pull it; otherwise, the measurement POD or guide rails may be damaged.
●
Installation of the measurement POD is done in the reverse procedure of removing it, as follows:
①
Apply both hands on both sides of the measurement POD and insert it in the POD installation opening of the main unit, and press it along the
guide rails to the bottom.
Never use excessive power to install it; otherwise, the measurement POD or guide rails may be damaged.
②
Install the measurement POD by clockwise rotating two clamp knobs on the front
Fig.1-1 POD clamp knobs
POD clamp knobs