B-H ANALYZER SY-8218
/ SY-8219
50
Chapter 7 Troubleshooting and Maintenance
■
7.1 Error messages and warning messages
●
If an error message or a warning message shown in Table 7-1 is displayed on the screen of this product, remedy described in the Table 7-1
should be made.
Table 7-1 Error / warning messages
Displayed messages
Causes
Remedies
E08
: Call service!
Interference of firmware when starting.
Restart according to
■
4.7.
When the symptom is the same even if it
restarts,
contact Iwatsu service center to tell
the error message No.
E01
,
E02
,
E03
,
E05
,
E07
,
E11
: Call service!
A malfunction occurs in hardware.
Restart this product.
When the symptom is the same even if it
restarts,
contact Iwatsu service center to tell
the error message No.
In case of SY-8218,
E12
: POD unconnected!
Shutdown
the SY-8218!
In case of SY-8219,
E12
: POD unconnected!
Shutdown
the SY-8219!
A measurement POD is not connected to this
product.
Power off this product and connect a
measurement POD to this product.
The model No. of a measurement POD
connected to this product is different from that
selected in measurement modes.
Power off this product and connect a proper
measurement POD.
If correspondence Option of Utilities >
Option is not Enabled, it is not possible to
measure.
E13
: Memory bank X data has NaN!
Non-numerical value is written in memory bank
X, or memory bank X is abnormal.
Restart this product.
When the symptom is the same even if it
restarts,
contact Iwatsu service center
to tell the error message No.
E14
:
Memory bank X shunt resistance errors!
The shunt resistance of memory bank X is
≦
0.
E16
:
SY-821x software version
is outdated!
As the version of the software for the B-H Analyzer
is old, it doesn't correspond to SY-810 used.
The main unit software should improve in the
version. Contact Iwatsu service center.
E17
: Temp. of FET heat sink on
DCS exceeds setting!
The temperature(thermistor) of the heat sink of
SY-961(DCS) exceeded 90
℃
.
Restart SY-961.
When the symptom is the same even if it
restarts,
contact Iwatsu service center to
tell the error message No.
E18
: DCS over current!
The output current of SY-961(DCS) exceeded
setting value ×1.1.
Power off SY-961.
Contact Iwatsu service center to tell the
error message No.
W09
:
Measurement was finished forcibly!
STOP button is pressed to stop measurement forcibly.
-
W12
: Temp. of the shunt resistance
exceeds setting!
The temperature of the shunt resistance
measuring the exciting current exceeds 80
℃
.
Reduce Hm, Bm, I
1
m, and V
2
m specified in
Fixed Parameter to decrease the exciting
current before measurement.
Reduce the number of Averages. Or increase
Tolerance to make the excitation time short.
W13
: Pod cover opens!
POD cover opens.
Close POD cover securely.
POD is removed from the main unit.
Power off this product and install POD on the
main unit securely.
W14
:
Shortage of the exciting current!
The exciting current is insufficient by
approximate value excitation.
Confirm whether the sample is connected.
Confirm whether the output has gone out of the
power amplifier.
Raise the gain of the power amplifier.
Increase N
1
of
the sample.
The DC cable (black) has been disconnected by
the DC bias measurement.
Confirm the disconnection of the DC cable
(black).
W15
: SST cover opens!
The cover of SY-956 (SST) is open.
Shut the cover of SY-956 neatly.
W16
: DCT cover opens!
The cover of SY-960 (DCT) is open.
Shut the cover of SY-960 neatly.