Instruction Manual 

 

 

 

 

-

 

Analyzer

 

SY-8218 / SY-8219 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание SY-8218

Страница 1: ...Instruction Manual Analyzer SY 8218 SY 8219...

Страница 2: ...2011_2017 IWATSU ELECTRIC CO LTD All rights reserved...

Страница 3: ...in death or serious injury Cautions Incorrect operation or failure to observe the caution may result in injury or damage to the instrument Notes Parts of the contents of this manual may be modified wi...

Страница 4: ...plifier optional The maximum voltage and current of the exciting power amplifier may be applied to the POD terminal which is connected to a sample To prevent danger put the provided POD cover on the P...

Страница 5: ...rt such as shaky base or inclined plane Dropping or falling down of this product could result in an electric shock injury or fire If this product is dropped or its cover is broken set the power switch...

Страница 6: ...cord Do not pull the power cord Do not forcibly bend the power cord Do not heat the power cord Do not twist the power cord Do not let the power cord get wet Do not bundle the power cord Do not put he...

Страница 7: ...erial or inflammable object through the ventilation port If any foreign object is put through the ventilation port an electric shock fire or failure may occur If any foreign object enters this product...

Страница 8: ...suitable for 100V system center voltage 100V to 120V is attached If the power supply voltage is 200V system center voltage 200V to 240V use 3 core power cord suitable for 200V system optional specifi...

Страница 9: ...or shock applied to this product during transportation may cause it to malfunction resulting in a fire The case where a failure or damage occurs when transporting this product with the measurement POD...

Страница 10: ...ributors Components B H Analyzer main unit 1 Standard Measurement POD 1 Accessories Power cord 1 Cord strap 1 POD cover 1 AC couple module 1 Power amplifier cable BNC BNC 1 OSC cable BNC SMA 1 Standar...

Страница 11: ...unit and accessories Accessories 1 Instruction manual CD Standard sample The measurement POD and POD cover Accessories 3 Power cord and the like Main unit Outer packing box Main unit pad Side pads Acc...

Страница 12: ...s responsible should be repaired without any cost When sending a product to be repaired clearly write the product name serial number in the label on the rear of this product and description of the fai...

Страница 13: ...Connection of equipments 22 4 6 Connection of sample 23 4 7 Powering on and off 24 Chapter 5 Setting Before Measurement 25 5 1 Setting of Unit 25 5 2 Setting of Information 26 5 3 Setting of Print 27...

Страница 14: ...t 59 8 7 External memory 59 8 8 Power supply 59 8 9 Mechanical 59 8 10 Environmental conditions 59 8 11 Certification Standards 59 8 12Accessories 60 8 13 Outside appearance diagram 60 Chapter 9 Measu...

Страница 15: ...like Since the number of pieces of obtained waveform data is 8 192 points cycle i e 16 times as many as the Iwatsu existing product it can measure coercive force residual magnetic flux density and th...

Страница 16: ...excessive moisture or dust Otherwise electric shock or fire may occur Do not place an object on this product If an object is placed on this product the cover may contact the internal circuit resultin...

Страница 17: ...oth hands on both sides of the measurement POD and pull the measurement POD to your side horizontally along the guide rails Never use excessive power to pull it otherwise the measurement POD or guide...

Страница 18: ...s product shall not be guaranteed and repaired Never connect a keyboard or a mouse to the USB port of this product i e operation is not guaranteed Never connect other than a USB memory and options spe...

Страница 19: ...mplifies the signal allowing the exciting current i1 t to flow through the primary turns Fig 2 1 Shape of sample The exciting signal i1 t is converted into a voltage Vs t by the shunt resistor with a...

Страница 20: ...unit Meaning Symbol Typical unit Meaning Measurement values Pcv W m3 Core loss per volume Bm T Max magnetic flux density Pcm W kg Core loss per mass V2m V Max induced voltage deg Phase angle Hm A m M...

Страница 21: ...calculated by the expression 2 17 2 15 Complex permeability j imaginary unit 0 2 f0 f0 measurement frequency 2 16 mode 2 17 2 18 2 19 2 20 T c dt t V t i T N N P 0 2 1 2 1 1 e c cv V P P e c cm W P P...

Страница 22: ...B H ANALYZER SY 8218 SY 8219 10 2 21 2 22 2 23 2 0 2 L R Z 0 2 0 2 2 0 2 2 0 2 2 10 7 5 3 log 20 V V V V THD e e L N A R 2 1 0 0...

Страница 23: ...when Main power switch on the rear is turned on it is effective Measurement POD It connects a sample to be measured the standard POD for SY 8218 SY 951 the standard POD for SY 8219 SY 955 OSC OUTOUT I...

Страница 24: ...ront and three on the rear Every USB port can be used They are assigned characters of driver in order of attachment timing first E next F G H and finally I in the same manner as a computer SYNC OUT It...

Страница 25: ...keys It displays number keys 0 9 character keys upper lower cases symbols space For a key on which several characters are written a displayed character changes each time it is pressed Ex Each time pre...

Страница 26: ...direction of the key Function knob On Popup Edit Box Clockwise rotation allows a current character displayed on the edition cursor to change as follows Character A B Y Z a b y z A Number 1 2 9 0 1 Cou...

Страница 27: ...election of SAVE RECALLallows Popup List to appear If pressed when Popup List is displayed Popup List is closed UTILITIES key It displays Utilities screen If pressed when Utilities screen is displayed...

Страница 28: ...ified output destination in a specified file format The file format and output destination are specified in Print frame on Utilities screen for details see 5 3 Setting of Print SELECT key If Reference...

Страница 29: ...Sec 4 4 V2m 200 V i1m 6A Max Output of power amplifier is exceeded Sample inductance L Applying turns to sample Sec 4 4 Decision of Hm Bm f N1 N2 Connecting this product to Meas POD Sec 4 5 Connecting...

Страница 30: ...length of magnetic path 4 1 Effective net core area 4 2 Effective volume 4 3 Mass 4 4 Fig 4 2 Toroidal core EI core For an EI core with the shape shown in Fig 4 3 use the calculation expression to ob...

Страница 31: ...Otherwise it may fail Moreover do not use the power amplifier exceeding the rated power Otherwise it may fail Next obtain the maximum exciting voltage V1m which is obtained by converting Max induced...

Страница 32: ...sample and then obtain Max exciting current i1m i1m is the same as I1m in Section 4 4 Turns of a sample or later using expression 4 13 4 13 If i1m exceeds 6 0 A this product cannot make measurement In...

Страница 33: ...ent How to apply turns If No of primary turns N1 is equal to No of secondary turns N2 the bifilar winding shown in Fig 4 5 is recommended It is made by making a twisted pair using primary and secondar...

Страница 34: ...fier and this product Connect OSC OUTPUT at the lower right on the front panel of the main unit with INPUT of the power amplifier using the provided OSC cable BNC SMA Connect them with an optional OSC...

Страница 35: ...S2 terminals Short circuit P1 and S1 terminals Fig 4 9 1 coil method It is recmmended that Fixed Parameter by 1 coil method is set in I1m and Hm specification Because the secondary turns is not given...

Страница 36: ...the LCD screen of this product the start progress bar screen appears for approx 1 min the bar disappears and then this product is ready to operate When the progress bar appears after turning on POWER...

Страница 37: ...be selected Press or to close Utilities screen Fig 5 1 Utilities screen If the other item is to be changed Utilities screen need not be closed Table 5 1 Setting of units Explanatory note Bold Factory...

Страница 38: ...to be changed and press Or rotate to change a number and press to determine the setting Table 5 2 shows description of each output information and selectable items Press or to close Utilities screen T...

Страница 39: ...to be set and press to determine the setting Table 5 3 shows description of each item and selectable items Press or to close Utilities screen Table 5 3 Setting of screen hardcopy Explanatory note Bold...

Страница 40: ...de is used only for service and not disclosed to a customer 8 Maint Retry Power of 23 25 Free 8 Mov Avg Odd in 1 111 Selects the moving average number applied to sample data of a measurement signal wa...

Страница 41: ...Description PowerAmp IE 1125 Select a power amplifier name to connect The power amplifier name is displayed on the measurement screen and saved when data is saved If Other is selected Edit Box just u...

Страница 42: ...s product recognizes connected measurement POD automatically If measurement POD connected with the selected measurement mode is not corresponding error E12 is displayed when the measurement begins Aft...

Страница 43: ...d the sample constant Excited Conditions Sets the measurement condition Mode Displays the current measurement mode Select Indicates whether the currently displayed graph is Current for current measure...

Страница 44: ...enter alphanumeric characters and press or move the edition cursor to the alphanumeric character to be changed press Or rotate to change the number Press to determine the entry and to close Popup Edit...

Страница 45: ...ngs 6 2 mode cannot select a square wave This product generates the symmetry square wave with Duty 50 Items Selectable items input rules Description Function Sine Sine wave Selects a type of excitatio...

Страница 46: ...netization is not left and the measured value is displayed If H is not within the range the signal output is changed so that H can approach to Hm while exciting the sample In this period the 2nd measu...

Страница 47: ...ssage Measuring disappears and the measurement result is displayed see Fig 6 4 To forcibly stop measurement after starting the measurement press If no measurement is made at that time the measurement...

Страница 48: ...s pressed again the displayed graph returns to B H graph Two time axis graphs are displayed Waveforms displayed in each graph are shown in Table 6 3 Table 6 3 Measurement waveform for time axis graph...

Страница 49: ...edition cursor is set automatically Items to be measured by each cursor are shown in Table 6 4 Table 6 4 Cursor types and measurement items 6 5 Even if Cross or Grad cursor is set for the time axis g...

Страница 50: ...he point on the cursor is displayed on Cursor Values Pushing allows the movement amount of the cursor to toggle between Coarse movement in unit of 20 points and Fine movement in unit of 1 point For Fi...

Страница 51: ...The rotation center of Grad cursor is the center of Cross cursor Therefore if the position of the rotation center of Grad cursor is to be changed change the cursor to Cross cursor and move the cursor...

Страница 52: ...vertical cursor is displayed the vertical cursor moves and the value of the point on the cursor is displayed on Cursor Values Pushing allows the movement amount of the cursor to toggle between Coarse...

Страница 53: ...elected the displayed graph is scaled down and if a smaller value is selected the displayed graph is scaled up see Fig 6 11 Scales to be set are shown in Table 6 5 Selection of Reset allows the displa...

Страница 54: ...lue larger than a current setting is selected the displayed graph is scaled down and if a smaller value is selected the displayed graph is scaled up see Fig 6 13 Scales to be set are shown in Table 6...

Страница 55: ...B H ANALYZER SY 8218 SY 8219 43 scale scale scale scale Fig 6 12 Time axis graph is reduced I is reduced Fig 6 13 B and I in Fig 6 12 is scaled down...

Страница 56: ...ge the measurement condition and make another measurement The measurement result is Current To display Reference press Each time it is pressed the measurement condition and value change as shown in Ta...

Страница 57: ...B H ANALYZER SY 8218 SY 8219 45 Fig 6 15 Reference function in time axis graph...

Страница 58: ...17 The folder and the file where things except the character that can be input to the name with ten keys are included cannot be displayed However only _ underscore can be displayed If output is made w...

Страница 59: ...7 move the edition cursor to the file display area see Fig 6 20 Press to select the route as an output destination Press to move the edition cursor to Meas_Data folder see Fig 6 21 Press to display th...

Страница 60: ...operation as is made Save operation is done so that the existing file is overwritten At this time W68 is displayed When Yes is selected the superscription is carried out When No are selected the super...

Страница 61: ...e save file If an altered file is read out resulting in wrong operation or failure this product shall not be guaranteed and repaired 6 9 Reference waveform Storage waveform cannot be saved In addition...

Страница 62: ...6 SY 821x software version is outdated As the version of the software for the B H Analyzer is old it doesn t correspond to SY 810 used The main unit software should improve in the version Contact Iwat...

Страница 63: ...and units of N1 and Le are proper Confirm that value of Hm specified in Fixed Parameter is proper Increase N1 of the sample W34 Setting condition error Yoke V3 or V4 200V It is forecast that the outp...

Страница 64: ...sensor of the heat sink of SY 961 DCS exceeded 90 Restart SY 961 When this phenomenon happens frequently contact Iwatsu service center to tell the error message No W58 Temp of sample on DCT exceeds se...

Страница 65: ...Do you want to overwrite it Data tries to be overwritten on an existing file Select Yes or No W69 Failed in a saved Data fails to be saved Memory is full or memory is removed during save operation Re...

Страница 66: ...cify Hm for Fixed Parameter If the number of turns changes a core loss value changes greatly The inductance of a sample resonates with the input capacitance of this product Reduce the inductance of th...

Страница 67: ...oistened with a small quantity of neutral detergent Never use organic solvent If used discoloration or corrosion may occur 7 5 Calibration Long time use may cause the measurement accuracy of this prod...

Страница 68: ...cal value f 100 kHz 50 mA 50 mV range or more at the amplitude of 80 or more of used range Amplitude 2 Typical value f 1 kHz 50 mA 50 mV range or more Core loss 5 6 Typical value estimated value for 8...

Страница 69: ...field a Relative permeability I1m A Max exciting current Bm T Max magnetic flux density a Relative permeability V2m V Max induced voltage L H Inductance Br T Residual magnetic flux density R Resistanc...

Страница 70: ...nd 2 Fixed Parameter target parameter any of B H I and V Setting of Equipment PowerAmp used power amplifier any of IE 1125 IE 1125A IE 1125B HSA4014 IW HSA4052 IW HSA4101 IW Other and Not Used However...

Страница 71: ...60 Hz Power consumption 130 VA Max Standby power consumption Approx 15 VA at 100 V approx 25 VA at 240 V only when main power SW is ON on the rear 8 9 Mechanical Outside dimensions 420 W 480 D 266 H t...

Страница 72: ...12 Accessories Power cord 2 5 m Un shielded Cord strap OSC cable BNC SMA1 m shielded Power amplifier cable BNC BNC 1 2 m shielded AC couple module Instruction manual CD POD cover User s guide 8 13 Ou...

Страница 73: ...hereinafter called high harmonic ingredient is included Fig 9 2 is the vector diagram indicating the case where the apparent power of the basic wave is VAand the apparent power of the high harmonic wa...

Страница 74: ...of current or voltage waveform and corresponds to the time gap of the measured waveform Here assume that real effective values of the current and voltage are i and V the real value of the phase angle...

Страница 75: ...urement accuracy of the core loss PC becomes worse than described in Section 9 2 This cannot be also escaped from the pure mathematical principle When both the primary exciting current i1 and the seco...

Страница 76: ...erseas Sales Sect Sales Dept No 2 Address 7 41 Kugayama 1 chome Suginami ku Tokyo 168 8511 Japan Phone 81 3 5370 5483 Facsimile 81 3 5370 5492 URL http www iti iwatsu co jp E mail info iti iwatsu co j...

Страница 77: ...SY 8218 SY 8219...