57
Cursor measurement in XY display
Time related measurement is not available in XY display.
Measurement type
Content
Operable cursor
CH
∆
V
Potential difference between H cursors
H-C1, H-C2
Potential difference between V cursors
V-C1, V-C2
Measurement example of CH
∆
V
←
Cursor that can be operated by FUNCTION knob
←
V measurement in vertical direction
(measured by CH2 volts/div)
←
V measurement in horizontal direction
(measured by CH1 volts/div)
←
Displays ******* if XY3&4 is not selected.
Allows observation of the phase difference between signals.
XY1&2
: Displays a waveform by expressing the horizontal direction
(X) on the CH1 voltage axis and the vertical direction (Y) on the CH2
voltage axis.
XY3&4
: Displays a waveform by expressing the horizontal direction
(X) on the CH3 voltage axis and the vertical direction (Y) on the CH4
voltage axis.
XY1&2/3&4
: Displays XY1&2 and XY3&4 simultaneously.
Measurement example of 90° phase different sine waves
XY display
XY1&2
YT display
XY1&2 display
Displays a nearly
complete round Lissajous
waveform.
90°
Limitation on XY display
• Calculated waveform(ADD/Sub/Mult) display, reference waveform display or
automatic measurement is not possible.
• Always displays with dots irrespective of Join setting.
• In the roll mode, waveforms are displayed in YT mode and if the state is changed from
RUN to STOP by pressing the RUN/STOP button, waveforms are displayed in XY mode.
Содержание DS-8814
Страница 18: ...xviii Memo...
Страница 19: ...Operation Manual...
Страница 22: ...3 Rear View GP IB port RS 232C port AC power inlet Fuse holder PCMCIA Type II slot Upper View Built in printer...
Страница 61: ...42 Memo...
Страница 100: ...81 Section 3 Daily Care...
Страница 106: ...87 Section 4 Specifications...
Страница 113: ...94 Memo...
Страница 114: ...F 1 Section 5 FFT Computation...
Страница 137: ...Memo...