Every capturing of an input signal averages data on the same time
axis and displays the waveform.
Selecting Average shows the Count on the menu. Set an average count
(2, 4, 8, 16, 32, 64, 128 and 256) depending on the noise condition.
The greater the count, the greater the effect of eliminating noise in the
signal becomes, but processing takes more time.
Example of measurement when averaging is not performed
Example of measurement when averaging is performed
Noise in the signal is eliminated by averaging.
Averaging restrictions
In the roll mode or long memory, Norm smpl (normal sampling) is set automatically
even if Average is selected.
Average
(Average)
A verage
count
(Count)
Sets average count to 64. L is effective.
Sets acquisition mode to Average.
Indicates averaging operation.
Averaging is not performed in the roll mode (short
memory with sweep time slower than 100 ms/div).
Averaging is not performed with long memory.
Performs averaging even when equivalent
sampling is set.
In the single sweep mode, averaging is
stopped after specified averaging cycles have
been performed completely.
Indicates normal sampling.
Sets acquisition mode to Norm Smpl.
Signal with large noise
Since a signal (sine wave of approx.
130 kHz) includes noise of over 10
MHz, the trigger coupling is set to
HF-R to stabilize triggering.
Содержание DS-8814
Страница 18: ...xviii Memo...
Страница 19: ...Operation Manual...
Страница 22: ...3 Rear View GP IB port RS 232C port AC power inlet Fuse holder PCMCIA Type II slot Upper View Built in printer...
Страница 61: ...42 Memo...
Страница 100: ...81 Section 3 Daily Care...
Страница 106: ...87 Section 4 Specifications...
Страница 113: ...94 Memo...
Страница 114: ...F 1 Section 5 FFT Computation...
Страница 137: ...Memo...