54
2. Start the calibration.
The display shows the last calibration value and the date of the
last calibration.
Press the START key on the membrane-sealed keypad to start the
calibration. Follow the instructions on the display.
5.4.3 Temperature calibration
The thermocouple and heating muffle of the furnace may be subject to changes which affect the furnace temperature,
depending on the mode of operation and the frequency of use. Conduct an automatic temperature calibration at least
every six months.
Temperature calibration requires only few steps:
1. Open the temperature calibration.
Scroll to page 2 in the home screen and press the
[Temperature
Calibration]
button.
The furnace must be heated up and have reached its
stand-by temperature (403 °C) before the calibration is
started.
3. Remove the firing plate.
Remove the firing plate from the furnace using the furnace tongs
and place it on the cooling tray.
5. Operation and Configuration
Vacuum hours
Number of operating hours of the vacuum pump
IP address
IP address indication
WLAN IP address
WLAN IP address indication
Internet connection
Indicates if the furnace is connected to the Internet
MAC address
Indication of the MAC address
WLAN MAC address
Indication of the WLAN MAC address
Press procedures
Number of all press cycles
Press cycles since last calibration
Number of press cycles since the last calibration