2. Safety First

This chapter is especially important for individuals who work with 
the Programat CS4 or who have to carry out maintenance or repair 
work. This chapter must be read and the corresponding instructions 
followed!

2.1   Indications

The Programat CS 4 has been especially designed for the sintering 
and glazing of ZrO

2

 materials such as IPS e.max ZirCAD as well as 

glazing and crystallization of IPS e.max CAD, for example. The  
Programat CS4 should be used for this purpose only. Other uses than 
the ones stipulated, e.g. cooking of food, firing of other materials 
etc., are contraindicated. The manufacturer does not assume any  
liability for damage resulting from misuse. The user is solely  
responsible for any risk resulting from failure to observe these 
Instructions.

Further instructions to assure proper use of the furnace:
–  The instructions, regulations and notes in these Operating  

Instructions must be observed. 

–  The instructions, regulations and notes in the material’s  

Instructions for Use must be observed.

–  The furnace must be operated under the indicated environmental 

and operating conditions (see chapter 9).

–  The Programat CS4 must be properly maintained.

 

 

Do not

 use colouring liquids containing chlorine or hydro- 

chloric acid in the Programat CS4 furnace. Due to the 
aggressive components of the liquids, the furnace surfaces 
or individual parts may be chemically corroded or damaged. 

If colouring liquids containing chlorine or hydrochloric acid are used, 
irritating gases may develop during the heating phase.

2.1.1

 

Contraindication   

Do not carry the furnace by the  
cooling tray.  

2.1.2

 

Contraindication 

Do not carry the furnace head by the 
cables, since the cables and  
connections may be damaged.

2.1.3

 

Contraindication 

The furnace head must not be 
removed from the furnace base as 
long as the furnace head is connected 
by means of the heater cable. 

2.1.4

Contraindication 

The furnace head is equipped with an 
electric drive and can be operated by 
means of the electronic controls. Never 
open the furnace head by hand, since 
this will damage the transmission.

2.1.5

 

Contraindication

Do not touch the thermocouple and 
the heating element in the sintering 
chamber. Avoid contact with the skin 
(grease contamination), as the parts 
may be prematurely damaged. 

Содержание Programat CS4

Страница 1: ...Programat CS4 Operating Instructions...

Страница 2: ......

Страница 3: ...5 5 Settable parameters and possible value ranges 5 6 Settings test programs and information 5 7 Explanation of the display symbols 5 8 Explanation of the acoustic signals 5 9 OSD 6 Practical Use 23 6...

Страница 4: ...nics plug 24 Leave spring 25 Air vents base 26 Cooling tray 27 Cooling tray fastening screw 28 Silicone washer 29 Hood 30 Attachment screw for hood 31 Air vents furnace head 32 Air vents rear panel 33...

Страница 5: ...5 44 16 26 4 8 27 28 15 25 1 3 6 5 37 7...

Страница 6: ...6 39 17 22 18 23 21 20 9 38 46 47 19 24 35 9 10 11 10 11 14 31 30 29 42 45 13 12 32 42...

Страница 7: ...key 81 Information key 82 Program 1 key 83 Program 2 key 84 Program 3 key 85 Program 4 key 86 Next program key 87 Open furnace head 88 Close furnace head 100 Sinter tray 102 Cleaning brush 103 IPS e m...

Страница 8: ...crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions Apparatus Programat CS4 Target group Dentists dental professionals These Operating Instructions facilit...

Страница 9: ...ust be operated under the indicated environmental and operating conditions see chapter 9 The Programat CS4 must be properly maintained Do not use colouring liquids containing chlorine or hydro chloric...

Страница 10: ...ace 2 1 10 Contraindication Do not insert any foreign objects into the air vents There is a risk of electric shock 2 1 11 Burn hazard Never place objects in the hot firing chamber by hand since there...

Страница 11: ...ks and dangers may perform these procedures After maintenance the required safety tests high voltage resistance protective conductor etc have to be carried out Ensure that only fuses of the indicated...

Страница 12: ...uctor Protection from electric shock Electric fuses Protection from electric shock Furnace housing and end caps Protection from electric shock burning and crushing 3 3 Description of functions The sin...

Страница 13: ...2 4 5 2 3 4 5 3 1 5 2 3 4 5 3 4 2 Selecting the location Place the furnace on a fire proof flat table using the rubber feet Make sure that the furnace is not placed in the immediate vicinity of heater...

Страница 14: ...strongly against the rear of the furnace For safety reasons the spacer must always be mounted Secure the cooling tray 26 with the two screws 27 including the silicone washers 28 Place the cooling tra...

Страница 15: ...er on the furnace head and carefully position it on the furnace head mounting 35 Ensure that the furnace head mounting mark 33 is aligned with the furnace base mounting mark 34 Make sure that the sint...

Страница 16: ...tablishing additional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 12 with the power socket...

Страница 17: ...1 Connect the power cord 12 with the wall socket 2 Put the On Off switch 9 at the rear of the furnace on position I Subsequently the furnace conducts an automatic self test The performance of all furn...

Страница 18: ...tarts the selected program Start LED 74 Lights up once a program has started It flashes when the program is paused Stop key 75 Pressing 1 x pause program Pressing 2 x program is aborted The STOP key a...

Страница 19: ...ive error number is indicated Value range C Value range F Symbol Parameters min max Unit min max Unit v1 Predrying stage 1 Predrying temperature 0 100 300 C 0 212 572 F V1 Predrying time 00 00 00 20 h...

Страница 20: ...red volume can be set using the keys Melody The desired melody may be set using the keys Programming Enables the programming of the parameters for the currently selected program Settings Display Short...

Страница 21: ...C or F Calibration value MHT Calibration value at 1500 C 2732 F The value is displayed in accordance with the selected temperature mode in C or F Information Display Short Description Mains voltage Sh...

Страница 22: ...ith the error tune endless beep The beeper may be switched off with the STOP key while the error message still remains visible If the error message is acknowledged with ESC the beeper also stops At th...

Страница 23: ...2 Loading for sintering procedure The sinter tray 100 holds up to approximately 6 units depending on the program Make sure that the objects do not touch in order to prevent them from sintering togeth...

Страница 24: ...icates a temperature below 400 C 752 F Step 2 Open the furnace head using the open furnace head key 87 Use a sinter tray fork to place the loaded sinter tray into the furnace Step 3 Press the Start ke...

Страница 25: ...a Settings Renaming the keyboard can be selected provided the currently selected program is not write protected The desired letters can be selected using the cursor keys circular The letter is selecte...

Страница 26: ...rticles see image a Fine glass dust The insulation of the lower part of the furnace and the cooling tray show residue of fine glass dust This type of contamination is more difficult to discern 7 1 2 C...

Страница 27: ...be cleaned from time to time What When Cleaning material Furnace housing 7 and furnace head if required soft dry cloth Membrane sealed keypad 8 weekly soft dry cloth Cooling tray 26 daily cleaning bru...

Страница 28: ...UTH 1 Press the Settings key 78 to access the Settings screen Select Temperature Calibration UTH using the cursor keys 79 80 The following display appears The UTH calibration program is selected with...

Страница 29: ...correction value can be transferred to the furnace with the help of the keys Each keystroke changes the furnace temperature by 1 C The correction value must be added to the current temperature of the...

Страница 30: ...oltage The furnace is operated with the incorrect supply voltage Make sure that the furnace is operated with the supply voltage indicated on the rating plate 1310 Calibration reminder Some time has pa...

Страница 31: ...ement bent or fractured Switch off the furnace and contact the Ivoclar Vivadent After Sales Service Damage of the thermocouple Is the thermocouple damaged or fractured Contact the Ivoclar Vivadent Aft...

Страница 32: ...kg Furnace head 17 kg Safety notes The sinter furnace complies with the following standards IEC 61010 1 2010 EN 61010 1 2010 UL 61010 1 2012 2015 CSA 61010 1 2012 2015 IEC 61010 2 010 2014 EN 61010 2...

Страница 33: ...contact the Ivoclar Vivadent Service Center Important information The current program tables are also available at www ivoclarvivadent com The program tables can be downloaded from the Internet as PDF...

Страница 34: ......

Страница 35: ......

Страница 36: ...ivadent S A de C V Calzada de Tlalpan 564 Col Moderna Del Benito Ju rez 03810 M xico D F M xico Tel 52 55 50 62 10 00 Fax 52 55 50 62 10 29 www ivoclarvivadent com mx Ivoclar Vivadent BV De Fruittuine...

Отзывы: