24
Step 4:
The furnace head opens automatically at the end of the program.
Note
The sintering chamber is still very hot.
Use the sinter tray fork to remove the sinter tray and place it on the
cooling tray. Allow the restoration to cool down to room
temperature before further processing.
You must not position a new restoration in the sintering
chamber before the display indicates a temperature below
700 °C /1292 °F.
6.4 Loading for glaze and crystallization programs
For glaze and crystallization programs, no more than 6 units can be
positioned in the centre of the crystallization tray. For more detailed
information, please refer to the instructions for use of the corre-
sponding material.
Depending on the program, choose either the IPS e.max
CAD crystallization tray or the IPS e.max Speed
crystallization tray. To stabilize the restoration on the
firing tray, only use IPS e.max CAD Crystallization Pin or
IPS Object Fix Putty/Flow. For more detailed information,
please refer to the instructions for use of the corresponding
material.
Use a sinter tray fork to place the crystallization tray in the furnace.
The crystallization tray must be positioned in the centre
of the furnace. Firing without a sinter tray is contra-
indicated.
6.5 Glazing and crystallizing procedure with a standard
program
Step 1:
Select the desired program using the program keys.
You must not position a restoration in the sintering
chamber before the display indicates a temperature below
400 °C /752 °F.
Step 2:
Open the furnace head using the open furnace head key (87). Use a
sinter tray fork to place the loaded sinter tray into the furnace.
Step 3:
Press the Start key (73) to start the program.
You can observe the cycle on the program progress indicator.
Step 4:
The furnace head opens automatically at the end of the program.
Note
The sintering chamber is still very hot.
Use a sinter tray fork to remove the sinter tray together with the
loaded crystallization tray and place it on the cooling tray. Allow the
restoration to cool down to room temperature before further
processing.
You must not position a new restoration in the sintering
chamber before the display indicates a temperature below
400 °C/752 °F.
6.6 Individual programs
Step 1:
Select a free program.
Step 2:
Select the parameter list via
Settings – Programming, and then
change the program parameters
using the -/+ keys. Then return to
the stand-by screen using ESC or a
program key.
Step 3:
Open the furnace head using the
open furnace head key (87)
and place the loaded sinter tray
(with or without crystallization
tray) into the furnace.
Press the Start key (73) to start the
program. You can observe the cycle on the program progress
indicator.
Step 4:
The furnace head opens automatically at the end of the program.
Note
The sintering chamber is still very hot.
Use the sinter tray fork to remove the sinter tray and place it on the
cooling tray. Allow the restoration to cool down to room tempera-
ture before further processing.
You must not position a new restoration in the sintering
chamber before the display indicates a temperature below
400 °C/752 °F.