3.1.8   Acceleration

The Acceleration setting determines how quickly the pump piston reaches the dispense/load rate from a stopped position. It
is also used to determine deceleration; how quickly the piston slows down when heading towards stop. Increasing the
acceleration may help increase the shearing of certain fluids from the tip improving repeatability. However, with higher
accelerations, high viscosity fluids may be more susceptible to causing the Actuator Module motor to stall.

Advantages

Disadvantages

Decreased Acceleration

Lower probability of stalls

Less shear at tip

Lower probability of fluid cavitations

Slightly longer cycle time

Increased Acceleration

More shear at tip

Higher probability of stalls

Slightly shorter cycle time

Higher probability of fluid cavitations

3.1.9   Drawback

Drawback is provided in some operating modes (Dispense, Dispense MCV, Meter) to allow the fluid to move in the
opposite direction after a dispense to better separate the fluid at the tip. Often, high viscosity or “stringy” fluids are
susceptible to poor separation at the end of a dispense. Drawback provides a snapping motion in the fluid to attempt
to shear it off at the tip. There are three parameters available for configuration of Drawback: Drawback Volume,
Drawback Rate, and Drawback Dwell.

The Drawback Volume determines the amount of fluid to move backwards after a dispense. Typically, this is
configured for a small fraction of the dispense volume.

The Drawback Rate determines the rate of fluid movement during the backwards drawback.

The Drawback Dwell determines the amount of time the fluid movement pauses between the end of the cycle and the
beginning of the next cycle. Typically, this is left at the default value, but may be increased if it benefits the application.

3.1.10  Torque

The torque setting determines the peak current available to the motor. Increasing the torque decreases the
probability of stalling with high viscosity fluids. Decreasing the torque increases efficiency by decreasing steady state
heat loss in the motor when using low viscosity fluids.

Advantages

Disadvantages

Decreased Torque

Reduced heat loss in the motor

Higher probability of stalls

Increased Torque

Lower probability of stalls

Increased heat loss in the motor

3.2

STANDARD OPERATION

This Standard Operation section provides the information and instructions for the most common operation of the
system. Most of the screens and the less common information have been removed. If you are an advanced user or
need additional information please refer to Section 3.3 - Advanced Operation.

The Controller Module provides the controls for producing fluid flow via a positive displacement pumping mechanism.
The systems utilize solid-state electronics, stepping motor drives, and precision machined ceramic pump heads.
These components combine to provide exceptional accuracy and precision, high reliability, and low maintenance.

Volume commands for the Controller Module use microliters. Rate commands are in microliters per second. Pumping

is started using the Start push-button

 based on the screen being viewed. Push the Stop push-button

 to

stop the operation before the respective volume setting is reached.

Содержание DIGISPENSE 3020

Страница 1: ...O Connector 3 6 3 1 2 7 MOTOR Connector 3 6 3 1 2 8 Power Entry Module 3 6 3 1 3 Fluid Movement 3 6 3 1 4 Dispense Load Rate 3 7 3 1 5 Direction 3 7 3 1 6 Active Port 3 7 3 1 7 Pump 90 Degree Offset 3...

Страница 2: ...tate Screen 3 49 3 3 8 5 Dispense Screen 3 51 3 3 8 6 Dispense Mult Screen 3 53 3 3 8 7 Dispense MCV Screen 3 55 3 3 8 8 Meter Screen 3 57 3 3 8 9 Meter Mult Screen 3 59 3 3 8 10 Setup A Screen 3 61 3...

Страница 3: ...Installation 3 111 3 5 Maintenance 3 111 3 5 1 Assembly Disassembly Procedures 3 111 3 5 1 1 Main Power Fuse 3 111 3 6 Problem Guide 3 112 3 7 Specifications 3 112 3 8 Model Number 3 112 3 9 Illustra...

Страница 4: ...terface for controlling the system The following items are located on the front panel 1 Function Push buttons 5 Stop Push button 2 Display 6 Idle Indicator 3 Start Push button 7 Arrow Push buttons 4 A...

Страница 5: ...sh buttons move the cursor on the display and are used to change values 3 1 1 8 Enter Push button Figure 3 1 Item 8 The Enter Push button has two functions the first function selects a value to change...

Страница 6: ...e 3 2 Item 6 The Logic I O connector is used for making the electrical connection to the LOGIC I O Interface Device Refer to section 3 3 10 3 3 1 2 7 MOTOR Connector Figure 3 2 Item 7 The Motor connec...

Страница 7: ...e fluidic characteristics etc This means the dispense rate is often lower than the load rate to increase repeatability and reduce the probability of fluid cavitations Advantages Disadvantages Decrease...

Страница 8: ...he bottom of the page show the output port locations for the four outlet positions In Forward Port A is always the input and Port B is the output In Reverse Port B is always the input and Port A is th...

Страница 9: ...awback Rate determines the rate of fluid movement during the backwards drawback The Drawback Dwell determines the amount of time the fluid movement pauses between the end of the cycle and the beginnin...

Страница 10: ...ed the operation will terminate and drawback will occur if enabled This means that a metered dispense may only occur up to volumes of a pump chamber volume minus the drawback volume If the Metering is...

Страница 11: ...be displayed Press the FPB1 key again to clear the help information from the screen 3 2 6 Permission Levels There are three levels of permission Operator I O Test and Supervisor Each level allows acc...

Страница 12: ...up Displays the current recipe number and pump size Recipe Used for saving a new recipe or retrieving an existing recipe Select New Value Used for selecting a new value from a list Setup A Displays r...

Страница 13: ...NSE CONTRAST 15 SCREEN ADJUST OR if prime mode and meter mode 14 PRIME METER CONTRAST 15 SCREEN SCREEN ADJUST OR if prime mode and disabled production mode 14 PRIME CONTRAST 15 SCREEN ADJUST OR if dis...

Страница 14: ...Start push button is not functional while this screen is visible Status The status indicates the operational state of the system Refer to section 3 3 9 for an explanation of each status type Recipe T...

Страница 15: ...5 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 DISPENSE SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE OR if direction change disabled 14 DISPENSE SETUP A GET 15 SCREEN SCREEN RECIP...

Страница 16: ...e volume of the pump and tubing The range is based on the pump size as shown in the Pump Minimum Maximum Chart The increment is equal to the minimum volume for the selected pump size Priming may be te...

Страница 17: ...d 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 PRIME SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE OR if load required 14 PRIME SETUP A LOAD GET 15...

Страница 18: ...mount of liquid dispensed in micro liters during a dispense cycle The range is based on the pump size as shown in the Pump Minimum Maximum Chart The increment is equal to the minimum volume for the se...

Страница 19: ...r pump selected 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 PRIME SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE OR if load required 14 PRIME SETUP...

Страница 20: ...se and Meter modes The direction is either Forward or Reverse Dispense Meter Rate The Dispense Meter Rate is the rate of the piston in micro liters per second during the meter operation The range is b...

Страница 21: ...abled fluidic setup mode 14 METER SETUP B SAVE 15 SCREEN SCREEN RECIPE OR if disabled production mode and disabled fluidic setup mode 14 GET SETUP B SAVE 15 RECIPE SCREEN RECIPE OR if prime mode and f...

Страница 22: ...section 3 3 1 for description of advanced modes Fluidic Setup Mode The Fluidic Setup Mode configures the current mode of operation during setup The mode can either be Disabled Prime Bubble Clr Clear o...

Страница 23: ...Display Change N Not Accessible Viewable Operator 0 DRAWBACK status 2 3 4 DRAWBACK VOLUME 5 DRAWBACK RATE 6 DRAWBACK DWELL 7 8 9 10 11 12 13 14 15 BACK Displays the Controller Module Status D D D Dis...

Страница 24: ...se volume plus drawback volume exceeds the full chamber volume Drawback Rate The Drawback Rate is the rate of the pump in micro liters per second during a drawback operation The range is based on the...

Страница 25: ...erator 0 SELECT NEW VALUE 2 3 Value Label 4 Selection 1 5 Selection 2 6 7 8 9 10 11 If help NOT pressed 12 13 If help pressed 12 Help message line 1 for selection 13 Help message line 2 for selection...

Страница 26: ...rameter to be changed and the current value Maximum Displays the maximum value for the parameter Minimum Displays the minimum value for the parameter Increment Displays the incremental value for the p...

Страница 27: ...ALID PASSWORD 6 7 8 9 CURRENT PERMISSION OPERATOR 10 11 If help NOT pressed 12 13 If help pressed 12 Help message line 1 13 Help message line 2 14 TO CHANGE 15 HELP OPERATOR PASSWORD BACK Display chan...

Страница 28: ...not functional but the Stop push button is functional while this screen is visible Enter Password This field is used to enter the password for the desired permission level Each permission level has a...

Страница 29: ...d 12 Help message line 1 13 Help message line 2 14 15 HELP CANCEL 3 2 7 10 Warning Recipe Exists Screen Supervisor D Display P Push button Accessible I O Test C Display Change N Not Accessible Viewabl...

Страница 30: ...as shown The Start push button is not functional but the Stop push button is functional while this screen is visible Warning Recipe Exists Screen Description The Warning Recipe Exists screen is inten...

Страница 31: ...from above 0 FAULT status 2 3 WARNING trying to move a seized piston 4 may break actuator coupling Please 5 remove pump before initiating move 6 7 8 9 10 MOVE PISTON MOVE PISTON REFERENCE 11 CLEAR FAU...

Страница 32: ...able as shown The Start push button is not functional but the Stop push button is functional while this screen is visible The Fault description lists the operational fault The error code is a numerica...

Страница 33: ...c Connect the outlet fitting and tubing 3c d Connect the gland outlet fitting and tubing 3d optional 3 2 9 Start up a Switch the 1 0 power switch to the 1 position b The display will illuminate and sh...

Страница 34: ...Port 40 Pitch Actuator Module Actuator Connector 2a The numbers in correspond to the text instructions 3b 3d 3a 3c 20 Pitch Actuator Module Actuator Connector 2a Inlet 3b 3d Port A 3a 3c Standard Mic...

Страница 35: ...wn go to SETUP A screen and change Production Mode to METER To start metering press the START button or supply a trigger to the rear panel The system will meter based on the Direction Load Threshold R...

Страница 36: ...single Dispense using an end of the pump chamber the end of the pump chamber used is dependent on Direction This mode is typically used when small repeatable volumes with minimum reload time are desi...

Страница 37: ...he next Production Mode operation but will not be automatically initiated In this case a manual initiation is required Every mode causes an automatic loading of the pump chamber after every Production...

Страница 38: ...between Fluidic Setup Modes may be achieved using recipes 3 3 2 1 Prime Mode Prime Mode is the most typically used Fluidic Setup Mode and is used to prime the fluidic tubing and components from the re...

Страница 39: ...d to perform one or more Waste operations in the desired Production Mode after an Agitate operation in order to properly setup the fluid for repeatable Dispenses 3 3 3 Port Parking PARK PORT Parking t...

Страница 40: ...arameters and is the typical use setting Recipes can be selected and loaded but not changed See serial interface k1 command for additional restrictions 3 3 7 Recipes A recipe is a collection of the op...

Страница 41: ...umerical value Fault Displays the current fault I O Test Displays trigger in input 2 CC trig in input 3 input 4 and RS232 Allows changing ready out fault out config 3 out config 4 out and auxiliary ou...

Страница 42: ...B4 GET RECIPE SCREEN SAVE RECIPE SCREEN AUTOTRIGGER SCREEN PROD DWELLS SCREEN CHANGE PERMISSION SCREEN I O TEST SCREEN FPB1 traverses up the Screen Ladder FPB2 traverses down the Screen Ladder FPB3 ha...

Страница 43: ...NSE CONTRAST 15 SCREEN ADJUST OR if prime mode and meter mode 14 PRIME METER CONTRAST 15 SCREEN SCREEN ADJUST OR if prime mode and disabled production mode 14 PRIME CONTRAST 15 SCREEN ADJUST OR if dis...

Страница 44: ...Start push button is not functional while this screen is visible Status The status indicates the operational state of the system Refer to section 3 3 9 for an explanation of each status type Recipe T...

Страница 45: ...5 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 DISPENSE SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE OR if direction change disabled 14 DISPENSE SETUP A GET 15 SCREEN SCREEN RECIP...

Страница 46: ...e volume of the pump and tubing The range is based on the pump size as shown in the Pump Minimum Maximum Chart The increment is equal to the minimum volume for the selected pump size Priming may be te...

Страница 47: ...duction mode 14 SETUP A GET 15 SCREEN RECIPE OR if NO actutator pump selected 14 DISPENSE SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 DISPENSE SETUP A REFERENCE GET...

Страница 48: ...either 4A 3A 2A 1A B C D or NONE Discharge Rate The Discharge Rate is the rate of the pump in micro liters per second during a bubble clear operation The range is based on the pump size as shown in t...

Страница 49: ...EEN RECIPE OR if NO actutator pump selected 14 DISPENSE SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 DISPENSE SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE O...

Страница 50: ...Pump Minimum Maximum Chart The increment is equal to 20 of the minimum rate for the selected pump size Intake Rate The Intake Rate is the rate the pump will operate during the load portion of an agita...

Страница 51: ...d 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 PRIME SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE OR if load required 14 PRIME SETUP A LOAD GET 15...

Страница 52: ...mount of liquid dispensed in micro liters during a dispense cycle The range is based on the pump size as shown in the Pump Minimum Maximum Chart The increment is equal to the minimum volume for the se...

Страница 53: ...ARK GET 15 SCREEN PORT RECIPE OR if NO actuator pump selected 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 PRIME SETUP A REFERENCE GET 15 SCREEN SCREEN REQUI...

Страница 54: ...a dispense cycle The range is based on the pump size as shown in the Pump Minimum Maximum Chart The increment is equal to the minimum volume for the selected pump size Dispense Meter Rate The Dispens...

Страница 55: ...position 14 SETUP A PARK GET 15 SCREEN PORT RECIPE OR if NO actuator pump selected 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 PRIME SETUP A REFERENCE GET 1...

Страница 56: ...ispense and Meter modes The direction is either Forward or Reverse Dispense Volume The Dispense Volume is the amount of liquid dispensed in micro liters during a dispense operation The range is based...

Страница 57: ...r pump selected 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if reference required 14 PRIME SETUP A REFERENCE GET 15 SCREEN SCREEN REQUIRED RECIPE OR if load required 14 PRIME SETUP...

Страница 58: ...se and Meter modes The direction is either Forward or Reverse Dispense Meter Rate The Dispense Meter Rate is the rate of the piston in micro liters per second during the meter operation The range is b...

Страница 59: ...SETUP A PARK GET 15 SCREEN PORT RECIPE OR if NO actuator pump selected 14 PRIME SETUP A SETUP C GET 15 SCREEN SCREEN SCREEN RECIPE OR if prime mode and reference required 14 PRIME SETUP A LOAD GET 15...

Страница 60: ...ONE Direction The Direction is the direction of liquid flow through the Pump Module for Dispense and Meter modes The direction is either Forward or Reverse Dispense Meter Rate The Dispense Meter Rate...

Страница 61: ...abled fluidic setup mode 14 METER SETUP B SAVE 15 SCREEN SCREEN RECIPE OR if disabled production mode and disabled fluidic setup mode 14 GET SETUP B SAVE 15 RECIPE SCREEN RECIPE OR if prime mode and f...

Страница 62: ...section 3 3 1 for description of advanced modes Fluidic Setup Mode The Fluidic Setup Mode configures the current mode of operation during setup The mode can either be Disabled Prime Bubble Clr Clear o...

Страница 63: ...AULT I O TEST 15 SCREEN SCREEN SCREEN SCREEN OR if permission less than I O Test 14 SETUP A SETUP C 15 SCREEN SCREEN Displays the Controller Module Status D D D Display change the auxiliary output D D...

Страница 64: ...led Busy Discharge Busy Prod Only Busy Moving Ready Idle Ready for Prod Ref Required Load Required or Port Park Refer to section 3 3 10 3 for signal functions Input CC The Input CC setting determines...

Страница 65: ...OR if NOT configured 14 SETUP B SYSTEM CLEAR 15 SCREEN INFO TOTALIZER OR if reference required 14 SETUP B SYSTEM REFERENCE CLEAR 15 SCREEN INFO REQUIRED TOTALIZER OR if permission less than superviso...

Страница 66: ...ort B Refer to section 3 1 6 for additional information Valving Speed The Valving Speed is a percentage of the rate the piston flat moves between ports The range is 1 to 100 with an increment value of...

Страница 67: ...Display Change N Not Accessible Viewable Operator 0 DRAWBACK status 2 3 4 DRAWBACK VOLUME 5 DRAWBACK RATE 6 DRAWBACK DWELL 7 8 9 10 11 12 13 14 15 BACK Displays the Controller Module Status D D D Dis...

Страница 68: ...se volume plus drawback volume exceeds the full chamber volume Drawback Rate The Drawback Rate is the rate of the pump in micro liters per second during a drawback operation The range is based on the...

Страница 69: ...ble Viewable Operator 0 AUTO TRIG status 2 3 AUTO RETRIGGER 4 RETRIGGER DWELL SEC 5 DISPENSE COUNT 6 RELOAD COUNT 7 8 9 10 11 12 13 14 15 BACK Displays the Controller Module Status D D D Display chang...

Страница 70: ...nt of time in seconds between a dispense and the next auto retrigger The range is 0 0 seconds to 300 0 seconds with an increment value of 0 01 Refer to section 3 3 1 6 Dispense Count The Dispense Coun...

Страница 71: ...on Accessible I O Test C Display Change N Not Accessible Viewable Operator 0 PROD DWELLS status 2 3 4 PRE OP DWELL SEC 5 POST OP DWELL SEC 6 7 8 9 10 11 12 13 14 15 BACK Displays the Controller Module...

Страница 72: ...in Production mode The range is 0 00 seconds to 2 55 seconds with an increment value of 0 01 Refer to section 3 3 1 6 Post Op Dwell The Post Op Dwell indicates the time in seconds from the end of a p...

Страница 73: ...P Push button Accessible I O Test C Display Change N Not Accessible Viewable Operator 0 ENCODER status 2 3 4 ENCODER 5 ENCODER LINES lines 6 7 8 9 10 11 12 13 14 15 BACK Displays the Controller Module...

Страница 74: ...ng can either be Disabled if no encoder option purchased on the Actuator Module or Enabled if encoder option purchased on the Actuator Module Encoder Lines The number of encoder lines is based on the...

Страница 75: ...11 12 http www ivek com 13 if NOT faulted 14 SETUP C CONTRAST 15 SCREEN ADJUST OR if faulted 14 SETUP C FAULT CONTRAST 15 SCREEN SCREEN ADJUST Displays the Controller Module Status D D D Displays the...

Страница 76: ...tion 3 3 9 for an explanation of each status type Firmware Version The Firmware Version displays the current version of software installed on the Controller Module Firmware CRC The Firmware CRC confir...

Страница 77: ...erator 0 SELECT NEW VALUE 2 3 Value Label 4 Selection 1 5 Selection 2 6 7 8 9 10 11 If help NOT pressed 12 13 If help pressed 12 Help message line 1 for selection 13 Help message line 2 for selection...

Страница 78: ...rameter to be changed and the current value Maximum Displays the maximum value for the parameter Minimum Displays the minimum value for the parameter Increment Displays the incremental value for the p...

Страница 79: ...VALID PASSWORD 6 7 8 9 CURRENT PERMISSION OPERATOR 10 11 If help NOT pressed 12 13 If help pressed 12 Help message line 1 13 Help message line 2 14 TO CHANGE 15 HELP OPERATOR PASSWORD BACK Display cha...

Страница 80: ...not functional but the Stop push button is functional while this screen is visible Enter Password This field is used to enter the password for the desired permission level Each permission level has a...

Страница 81: ...SSWORD ENTER NEW VALUE conditionally visible 6 7 MAXIMUM 8 MINIMUM 9 10 INCREMENT 11 If help NOT pressed 12 13 If help pressed 12 Help message line 1 13 Help message line 2 14 15 HELP MINIMUM MAXIMUM...

Страница 82: ...imum Displays the maximum value for the password The maximum value is 65535 Increment Displays the incremental value for the password The increment value is 1 To change the password 1 Use the up and d...

Страница 83: ...y P Push button Accessible I O Test C Display Change N Not Accessible Viewable Operator 0 CONTRAST ADJUST 2 3 Press UP Arrow to increase contrast 4 Press DOWN Arrow to increase contrast 5 6 Press SELE...

Страница 84: ...permission level determines which push buttons are available as shown NOTE The Contrast Adjust screen may not be available on all units The Start push button is not functional but the Stop push button...

Страница 85: ...sage line 1 13 Help message line 2 14 15 HELP CANCEL 3 3 8 23 Warning Recipe Exists Screen Supervisor D Display P Push button Accessible I O Test C Display Change N Not Accessible Viewable Operator 0...

Страница 86: ...as shown The Start push button is not functional but the Stop push button is functional while this screen is visible Warning Recipe Exists Screen Description The Warning Recipe Exists screen is inten...

Страница 87: ...ton Accessible I O Test C Display Change N Not Accessible Viewable Operator 0 RESET PARAMETERS 2 WARNING 3 4 5 6 PUSH ERASE TO ERASE ALL RECIPES 7 AND RESET PASSWORDS 8 9 PUSH KEEP TO KEEP RECIPES AND...

Страница 88: ...l while this screen is visible Reset Parameter Warning Screen Description The Reset Parameter Warning screen is intended to prevent accidently resetting all the Controller Module parameters This scree...

Страница 89: ...from above 0 FAULT status 2 3 WARNING trying to move a seized piston 4 may break actuator coupling Please 5 remove pump before initiating move 6 7 8 9 10 MOVE PISTON MOVE PISTON REFERENCE 11 CLEAR FAU...

Страница 90: ...able as shown The Start push button is not functional but the Stop push button is functional while this screen is visible The Fault description lists the operational fault The error code is a numerica...

Страница 91: ...T 3 OFF 9 CONFIG OUT 4 OFF 10 AUXILIARY OUT OFF 11 RS232 RS485 12 IN xxxxxxxxxxxxxxxx disabled 13 OUT 232 test 14 15 BACK Displays the trigger in and input 3 setting N D D Displays the input 2 and inp...

Страница 92: ...r OFF if no external signal is applied or ON if an external signal is applied Ready Out Fault Out Config Out 3 Config Out 4 Auxiliary Out The status of the output signals are displayed and changed her...

Страница 93: ...perations ISOLATING The Isolation portion of an Agitate operation is active AGITATING The Agitate portion of an Agitate operation is active RETURNING The Return portion of an Agitate operation is acti...

Страница 94: ...g Config Serial Interface h12 h13 h14 X X X CLEAR FAULT 0 Activate input to clear fault and initiate reference if REF required X X X FLUIDIC MODE 1 Activate input to initiate the currently configured...

Страница 95: ...Reference Required C Idle Load Required C Idle Production Mode Disabled C Idle Fluidic Mode Disabled C Idle Not Configured C Idle Motion Stopped C Initializing C Faulted NC Referencing C Porting C Loa...

Страница 96: ...the 4th output Signal Levels All signals are optically isolated The power for all signals is provided by the customer s equipment All inputs accept a 24 VDC signal and require 20 mA All outputs condu...

Страница 97: ...not broadcast messages such as fault conditions but only responds when it receives a command Command Structure The command is a string of ASCII characters The use of the ASCII backspace or rub out ch...

Страница 98: ...s in the command string which are not required by the command specified will be ignored Following is an example of a command string p1 100 CR Command p1 valving speed a setting of 100 Transmission sho...

Страница 99: ...lue2 a3 value2 Returns current Autotrigger Count setting Autotrigger Count value2 2 minimum default a4 a4 value2 65535 maximum a4 value2 a4 value2 Returns current Autoreload Count setting Autoreload C...

Страница 100: ...ing h0 value2 h0 value2 Sets the Auxiliary Output configuration value2 0 Disabled 1 Conducts during fluid mode 2 Conducts during both default 3 Conducts during prod mode h1 h1 value2 Returns the Contr...

Страница 101: ...ady out idle 5 Ready Out for production operation default h23 h23 value2 Returns the Output 3 setting h23 value2 h23 value2 Sets the Output 3 configuration value2 0 Disabled 1 Busy during discharge on...

Страница 102: ...Permission value2 Password for Desired Permission Level in command Current Permission Level in response Maximum 65 535 Minimum 10 k3 k3 value2 Returns the current Power Up Permission setting k3 value...

Страница 103: ...percentage setting p1 value2 p1 value2 Sets the Valving Speed percentage value2 Maximum 100 default Minimum 1 p2 p2 value2 Returns the Park Position p2 value2 p2 value2 Sets the Park Position Maximum...

Страница 104: ...load rate r3 value2 r3 value2 Sets the current Fluidic Mode load rate in microliters per second value2 Maximum Pump Dependent Minimum Pump Dependent s STATUS s0 s0 value2 Returns the volume remaining...

Страница 105: ...mode v2 v2 value2 Returns the Load Threshold v2 value2 v2 value2 Sets the Load Threshold value2 Dependent on pump selection and Mode v3 v3 value2 Returns the Fluidic Mode isolation volume v3 value2 v3...

Страница 106: ...0 value2 y10 value2 Sets the Pump 90 degree offset value2 0 Disabled default 1 Enabled z SOFTWARE VERSION z or z0 z0 value2 Returns the Software Version as text This is the only command that returns a...

Страница 107: ...the affected channel is performed Faults cause the Fault screen section 3 2 7 11 to appear If a fault occurs the type of fault will be displayed on the screen Once the problem is corrected press the R...

Страница 108: ...or slightly higher than the Pump Module and the Pump Module even with or slightly higher than the dispense tip Adjust as necessary to fit your application a Connect the inlet fitting and tubing 3a b C...

Страница 109: ...e Port 40 Pitch Actuator Module Actuator Connector 2a The numbers in correspond to the text instructions 3b 3d 3a 3c 20 Pitch Actuator Module Actuator Connector 2a Inlet 3b 3d Port A 3a 3c Standard Mi...

Страница 110: ...id displaced the actual forward motion is the sum of the specified dispense volume and the drawback volume The Drawback Rate during drawback and the Drawback Dwell time between the forward and reverse...

Страница 111: ...ger to the rear panel The system will meter based on the Direction Rate and Load Rate Pressing the STOP push button can stop the priming any time 3 4 INSTALLATION General operating practices provide t...

Страница 112: ...O Trigger In Requirements 24 VDC 20mA max Customer power source AUX OUT Signal Requirements Maximum external voltage 48 VDC Maximum current 250 mA 3 8 MODEL NUMBER The model number provides important...

Страница 113: ...r Controller power on and opera Motor Cable Check the cable connection between the Controller tional but will not actuate pump Module and Actuator Module Inspect and repair motor faulty cable Cannot r...

Страница 114: ...up screen Press FPB4 to enter Contrast Adjust screen and then press Down Arrow until contrast is acceptable No fluid movement Pump 90 Degree Offset Make sure pump orientation matches this setting doe...

Страница 115: ...pense Rate Load Rate or Acceleration rate of piston movement settings Fluid viscosity too high for Increase Torque setting torque setting Tubing too small for rate of Increase tubing diameter fluid mo...

Страница 116: ...he recipe Unable to Erase All Recipes An error occurred when Cycle power and attempt to reset parameters resetting parameters again Unable to Erase At Least One An error occurred when Cycle power and...

Страница 117: ...ill reset all param corrupted eters and recipes Configuration Value Out of A stored configuration data Cycle power Bounds Fault is corrupted Reset parameters WARNING will reset all param eters and rec...

Отзывы: