73
Setting the Instrument Date and Time when the POCT-1A Communication Protocol is Used:
1.
Start HEMOCHRON
Configuration Manager
(see page 49).
2.
Read the configuration from the instrument that is connected over the network (see page 52).
Note:
When the POCT-1A communication protocol is used, HEMOCHRON Configuration
Manager cannot communicate with the instrument over the COM port.
3.
Click the
Clone Instrument
button on the Configuration Module dialog box (page 53) to display the
Clear Instrument Counts dialog box:
4.
Select
Synchronize Instrument Date + Time with PC (POCT-1A)
. Do not select the other two
options.
5.
Select
OK
. The Write Configuration dialog box is displayed (page 67).
6.
Click
Start
. Progress information is displayed in the status line.
7.
A message box is displayed when writing of the configuration data is complete. Click
OK
to
acknowledge the message.
Purging Configuration Data
An existing configuration file can be removed from the
Configuration Manager
database. A record of the file
removal is maintained in the audit trail.
1.
Select
Purge Configuration Data
from the
Utilities
menu. The Purge Configuration Data dialog
box is displayed:
2.
Use the drop-down arrow to select the configuration file that is to be purged. The information for the
configuration file is displayed.
Note:
The information is read-only and cannot be changed by the user.
3.
Click
OK
. The selected configuration file is purged from the database.
Rebuilding the Database
1.
Select
Rebuild Database
from the
Utilities
menu. The Rebuild Database dialog box is displayed: