57
3.
Pre-stored only
: Enter a check mark to specify the cuvette information can only be selected from a
list of lot numbers specified by the supervisor.
Note:
Up to five cuvette lot numbers and expiration dates can be specified for each assay (see
page 64).
Note
: If more than five cuvette lot numbers per assay are scanned or entered, the lot number(s)
with the closest expiration date will be deleted.
QC Lot Options
1.
Required
: Enter a check mark to require entry of a QC lot number when running a quality control
test.
2.
Scan only
: Enter a check mark to specify the QC lot number can only be entered using the barcode
scanner.
Note:
If
Scan only
is enabled, the operator is prompted to scan the QC information when
running a test. The operator can still enter the QC information manually by pressing the
7
key
(as indicated by the
7=Manual Entry
prompt), but the QC information must be entered twice.
Refer to page 33 for additional information.
3.
Pre-stored only
: Enter a check mark to specify the QC lot number can only be selected from a stored
list.
Note:
Up to five QC lot numbers, expiration dates, and ranges can be specified for each assay
for both level 1 and level 2 controls (see page 65).
Note
: If more than five QC lot numbers per assay are scanned or entered for level 1 and/or level
2 controls, the lot number(s) with the closest expiration date will be deleted.
Saving the Options
1.
Click
Save & Exit
to save the specified options and cancel display of the dialog box.
Note
:
Cancel
clears all fields without saving any parameters.