is intended for use in the electromagnetic environment specified below; The customer or the user of the
WP300
should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
8 kV contact
2 kV, +4 kV, + 8 kV,
+15 kV air
8 kV contact
2 kV, +4 kV, + 8 kV,
+15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical public low-voltage power
supply network that supplies buildings used for
domestic purposes, commercial or hospital, clinic
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.