Operation Manual
2.4.3
Patient Test
When the patient (and accompanying individual if needed) turns on the WatchPAT™
device by pushing the On/Enter key (round center button) for about 2 seconds a self-
diagnostic test is automatically performed and the following screen is displayed:
PATIENT 22:51
Please wait
Testing…
If the WatchPAT™ device passes this self-diagnostic test, the following screen will be
displayed:
PATIENT 22:51
GOOD NIGHT!!!
Time elapsed=9:50
Recording…
Note
During recording the display turns off to conserve battery life. Pressing
the On/Enter key (center button) during recording will turn on the display
for a few seconds.
If the WatchPAT™ device fails this self-diagnostic test, the following screen will be
displayed:
PATIENT 22:51
Error code=xx
Device S/N=xxxxx
Call Help Desk
TEST ABORTED