©Isothermal Technology
Page 53 of 61
914 millisKanner manual Iss. 1.00
Example: after discovering whether there are any millisKanners attached at power-up,
the milliK will send the command
MILL:INIT 10
to start numbering input channels on the
millisKanners from channel 10 upwards.
8.2.6.10
8.2.6.10
8.2.6.10
8.2.6.10
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:STARt?
skanner:STARt?
skanner:STARt?
skanner:STARt?
Format: MILLiskanner:STARt?
Reports the channel number of the first input (on the millisKanner to which the RS232
connection is made).
Example: in a system with two millisKanners connected to a milliK, the response to the
command
MILLI:START?
sent to the second (last) millisKanner in the daisy-chain is “20”.
8.2.6.11
8.2.6.11
8.2.6.11
8.2.6.11
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:CLOSe
skanner:CLOSe
skanner:CLOSe
skanner:CLOSe
Format: MILLiskanner:CLOSe<channel#>
{POSitive|NEGative}
This command is reserved for use by the milliK and should not be used directly. It is used
to select (connect input to output) the specified channel in the daisy-chain. All other
channels are automatically deselected. The polarity of the sense terminals is defined by
the
{POSitive|NEGative}
parameter. The millisKanner returns a <CR> (carriage return) to
acknowledge that the channel has been selected and is ready to use.
Example: to select channel 17 with normal (positive) polarity, use:
MILL:CLOS17 POS
8.2.6.12
8.2.6.12
8.2.6.12
8.2.6.12
Command: MI
Command: MI
Command: MI
Command: MILLi
LLi
LLi
LLiskanner:OPEN
skanner:OPEN
skanner:OPEN
skanner:OPEN
Format: MILLiskanner:OPEN<channel#>
This command is reserved for use by the milliK and should not be used directly. It is used