© Isothermal Technology
Page 96 of 102
923 milliK manual - issue: 1.10
Example: If the milliK had been calibrated by Joe Bloggs of Metrosol in their secondary
laboratory at 20°C ±2°C and had not had any calibration adjustment applied since it might
respond to
CAL:IDEN?
with: “JBloggs,Metrosol,Main_Lab,18-22C”
8.6.39
Command CALibrate:PASSword
Format: CALibrate:PASSword
<old password>,<password>,<password>
Changes the password used to lock (protect) the calibration of the milliK’s measurement
system. This is initially set to “1234”, but should be changed before using the milliK for any
critical or traceable measurement or calibration work. The new password must be at least 4
characters in length and must be typed in identically twice in order to effect the change.
Example: to change the password from 1234 to ABCD use:
CAL:PASS 1234,ABCD,ABCD
8.6.40
Command CALibrate:VALid?
Format: CALibrate:VALid?
Reports whether the calibration information loaded from the milliK’s database is valid. If it is
not valid it will provide an error code that should be reported to Isotech when requesting
technical support.
Example: a correctly calibrated milliK will respond to
CAL:VAL?
with: “Calibration Okay”